PCD Pump 2018-06-11 WebEx: Difference between revisions
Jump to navigation
Jump to search
| Line 48: | Line 48: | ||
:F2F discussion | :F2F discussion | ||
::New limits would further constrain the existing limits, would only apply to dose limits, not concentration, times etc, | ::New limits would further constrain the existing limits, would only apply to dose limits, not concentration, times etc, | ||
| Line 56: | Line 54: | ||
::CommonID - RxNorm, NDC code, GUID | ::CommonID - RxNorm, NDC code, GUID | ||
::Name – Common Drug Name | ::Name – Common Drug Name | ||
::Concentration | ::Concentration - unreduced drug amount and volume | ||
::Advisory | ::Advisory | ||
::Care Are/Unit/Ward | |||
:Potential Options: | :Potential Options: | ||
Revision as of 10:28, 11 June 2018
Meeting Purpose
IHE PCD Infusion Pump Working Group discussions.
WebEx Information
Topic: IHE PCD Infusion Pump WG
Date: Monday, June 11, 2018
Time: 11:00, Eastern Time (GMT -05:00, New York)
Duration: 60 Minutes
Note: Specific web & phone information will be provided via e-mail to group members.
Contact Kurt Elliason for more information.
Proposed Agenda
- 1. Review today's Agenda
- 2. Review 2018.05.14 WebEx Discussion Notes & Action Items
- 3. Review PIV Order limits from EMR
- 4. Review first pass of common data terms for library synchronization.
- 5. Review PCD Device Infusion Pump Action Items
- 6. New Business / Next Meetings
Attachments / Materials
- <add attachments links here>
- Purpose: EMR can send limits specific to an individual patient order which can override pump drug library limits
- CP needed: add descriptive text to PIV in TF to include order limits as optional functionality 2 use cases:
- • Limits entered by provider at order entry
- • Limits from drug or order configuration in EMR
- • 0 to 4 limits may be present in a PCD-03 message (0-1 of each term)
- New terms:
- MDCX_ORDER_LOWER_HARD_LIMIT
- MDCX_ORDER_LOWER_SOFT_LIMIT
- MDCX_ORDER_UPPER_SOFT_LIMIT
- MDCX_ORDER_UPPER_HARD_LIMIT
- Example:
- OBX|1|NM|9999^MDCX_ORDER_UPPER_HARD_LIMIT^MDC|1.1.1.1.|25|8888^MDC_blob_per_kg^MDC
- F2F discussion
- New limits would further constrain the existing limits, would only apply to dose limits, not concentration, times etc,
- Basic Info to Share between EHRs and Library Software:
- DrugRecordID – GUID (each Name, Concentration, Advisory combiniation)
- CommonID - RxNorm, NDC code, GUID
- Name – Common Drug Name
- Concentration - unreduced drug amount and volume
- Advisory
- Care Are/Unit/Ward
- Potential Options:
- InfusionSetupID – GUID (each protocol that references a DrugRecordID would have one of these)
- DrugRecordID
- Rate/Dose
- DosingUnits
Discussion Notes
- Chair/Host: Kurt Elliason (Smiths Medical)
- Participants: <tbd>
- NOTE: Action Item Review updates have been posted directly to the Infusion Pump Action Items wiki page
Item Topic Discussion 1 Agenda & Discussion Notes Review
- ChairStatus/Discussion:
Decisions/Issues:
Action(s):2 PIV Order limits from EMR
- tbdStatus/Discussion: Decisions/Issues:
Action(s):
3 Common library terms for synchronization
- tbdStatus/Discussion:
Decisions/Issues:
Action(s):
4 Next Meeting
- ChairStatus/Discussion: Decisions/Issues:
- Topics will include ...
Action(s):
Next Meeting
NOTE: Next Meeting will be held during the regularly scheduled time: 11:00 (Eastern) on June 25 (60 minutes)
(Reviewed & approved by PCD Infusion Pump WG on <TBD>