PCD Pump 2025-06-09 Teams: Difference between revisions
Jump to navigation
Jump to search
| (One intermediate revision by the same user not shown) | |||
| Line 45: | Line 45: | ||
:'''Chair/Host:''' ''Kurt Elliason (ICU Medical)'' | :'''Chair/Host:''' ''Kurt Elliason (ICU Medical)'' | ||
:'''Participants:''' '' | :'''Participants:''' ''Amy Rowe (BBraun), Brian Sullivan(BD), Evan O'Grady(Fresenius-Kabi), Javier Becerril (BD), Jeffery Gaetano (BD), Michael Workman (BD), Paul Schluter (Paul Schluter Consulting), Silas Morris (Epic), Tom Kowalczyk (BBraun) | ||
| Line 73: | Line 73: | ||
| '''Status/Discussion:''' | | '''Status/Discussion:''' | ||
:3a. Multi-step/bolus order examples | |||
:: CP-DEV-021 - ready for ballot, added MDC terms for steps to IPEC example, defer DEC examples to a future CP. | |||
:3b. MDC for Volume on screen | |||
:: Reviewed latest proposal. most pumps can't report what is displayed, only what has been delivered. Many MDC terms exist for reporting various totals, total for infusion, for container, for method, for period, for segment etc. Look into improving documentation for when each is used and get more consistency on use of existing terms. | |||
:3c. PCD-10 example | |||
::CP-DEV-023 - ready for ballot, added MDC terms for steps to IPEC example, will include time/date fields in OBX segments | |||
:3d. Rack/Stack clarifications and examples | |||
:: CP-DEV-020 - almost ready for ballot, clean up descriptions to remove NA and use empty fields, check with Michael F. Nicolas C. about NIST Connectathon tool error when using 2^^7. Should we use a different data type instead of ST. | |||
:4a. PCD-03 Error codes | |||
:: No update | |||
:4b. White paper updates | |||
:: no update | |||
:4c. New terms - drug not in library and completion mode for switching sources | |||
:: Add to agenda for next meeting. | |||
:: Drug not in library - terms exist for library on pump and drug in current source channel but nothing to identify if that drug/program is in the current library or not. | |||
:New topic | |||
::CP-DEV-022 -remove text "(not valid for PCA)" from XO in ORC-1 section. | |||
Latest revision as of 13:39, 9 June 2025
Meeting Purpose
IHE PCD Infusion Pump Working Group discussions.
WebEx Information
Topic: IHE PCD Infusion Pump WG
Date: Monday, June 9, 2025
Time: 12:00, Eastern Time (GMT -05:00, New York)
Duration: 60 Minutes
Note: Specific web & phone information will be provided via e-mail to group members.
Contact Kurt Elliason for more information.
Proposed Agenda
- 1. Review today's Agenda
- 2. Review 2025-05-12 Pump Teams Session
- 3a. Multi-step/bolus order examples Multi-step final draft CP
- 3b. PCD-10 example Final draft CP
- 3d. Rack/Stack clarifications and examples Final draft CP
- 4a. PCD-03 Error codes
- 4b. White paper updates
- 4c. New terms - drug not in library and completion mode for switching sources
- 5. Review PCD Device Infusion Pump Action Items
- 6. New Business / Next Meetings
Attachments / Materials
- <add attachments links here>
Discussion Notes
- Chair/Host: Kurt Elliason (ICU Medical)
- Participants: Amy Rowe (BBraun), Brian Sullivan(BD), Evan O'Grady(Fresenius-Kabi), Javier Becerril (BD), Jeffery Gaetano (BD), Michael Workman (BD), Paul Schluter (Paul Schluter Consulting), Silas Morris (Epic), Tom Kowalczyk (BBraun)
- NOTE: Action Item Review updates have been posted directly to the Infusion Pump Action Items wiki page
Item Topic Discussion 1 Agenda & Discussion Notes Review
- ChairStatus/Discussion:
Decisions/Issues:
Action(s):2 Agenda topics
- tbdStatus/Discussion: - 3a. Multi-step/bolus order examples
- CP-DEV-021 - ready for ballot, added MDC terms for steps to IPEC example, defer DEC examples to a future CP.
- 3b. MDC for Volume on screen
- Reviewed latest proposal. most pumps can't report what is displayed, only what has been delivered. Many MDC terms exist for reporting various totals, total for infusion, for container, for method, for period, for segment etc. Look into improving documentation for when each is used and get more consistency on use of existing terms.
- 3c. PCD-10 example
- CP-DEV-023 - ready for ballot, added MDC terms for steps to IPEC example, will include time/date fields in OBX segments
- 3d. Rack/Stack clarifications and examples
- CP-DEV-020 - almost ready for ballot, clean up descriptions to remove NA and use empty fields, check with Michael F. Nicolas C. about NIST Connectathon tool error when using 2^^7. Should we use a different data type instead of ST.
- 4a. PCD-03 Error codes
- No update
- 4b. White paper updates
- no update
- 4c. New terms - drug not in library and completion mode for switching sources
- Add to agenda for next meeting.
- Drug not in library - terms exist for library on pump and drug in current source channel but nothing to identify if that drug/program is in the current library or not.
- New topic
- CP-DEV-022 -remove text "(not valid for PCA)" from XO in ORC-1 section.
Decisions/Issues:
Action(s):
3 TBD
- tbdStatus/Discussion:
Decisions/Issues:
Action(s):
4 Next Meeting
- ChairStatus/Discussion: Decisions/Issues:
- Topics will include ...
Action(s):
- 3a. Multi-step/bolus order examples
Next Meeting
NOTE: Next Meeting will be held during the regularly scheduled time: 12:00 (Eastern) on June 23 (60 minutes)
(Reviewed & approved by PCD Infusion Pump WG on <TBD>