Proposed Agenda
- 1. Review today's Agenda
- 2. Review 2025-06-09 Pump Teams Session
- 3a. Multi-step/bolus order examples Multi-step final draft CP
- 3b. PCD-10 example Final draft CP
- 3d. Rack/Stack clarifications and examples Final draft CP
- 4a. PCD-03 Error codes
- 4b. White paper updates
- 4c. New terms - drug not in library and completion mode for switching sources
- 5. Review PCD Device Infusion Pump Action Items
- 6. New Business / Next Meetings
Attachments / Materials
- <add attachments links here>
Discussion Notes
- Chair/Host: Kurt Elliason (ICU Medical)
- Participants: Al Engelbert (BBraun), Amy Rowe (BBraun), Brian Witkowski (Baxter), Dan Lee (ICU Medical), Javier Becerril (BD), Jeffery Gaetano (BD), Lisa Digget (BD), Paul Schluter (Paul Schluter Consulting), Silas Morris (Epic), Steve Dain(IEEE), Tom Kowalczyk (BBraun)
- NOTE: Action Item Review updates have been posted directly to the Infusion Pump Action Items wiki page
Item Topic Discussion 1 Agenda & Discussion Notes Review
- ChairStatus/Discussion:
Decisions/Issues:
Action(s):2 Agenda topics
- tbdStatus/Discussion: - 3a. Multi-step/bolus order examples
- CP-DEV-021 - ready for ballot
- 3b. PCD-10 example
- CP-DEV-023 - ready for ballot
- 3c. Rack/Stack clarifications and examples
- CP-DEV-020 - not ready for ballot, checked with Michael F. & Nicolas C. about NIST Connectathon tool error when using 2^^7.
- OBX-5 is a repeatable element so ~ are allowed here.
- And since you are using ST, a primitive, ^ will not make sense and are not allowed.
- Should we use a different data type instead of ST. or use ST and not use empty strings with ~ (2,NA,NA,7) is 2~NA~NA~7 instead of 2^^^7
- CP-DEV-020 - not ready for ballot, checked with Michael F. & Nicolas C. about NIST Connectathon tool error when using 2^^7.
- 4a. PCD-03 Error codes
- no update today
- 4b. White paper updates
- no update today
- 4c. New terms - drug not in library and completion mode for switching sources
- Consider possible terms, MDCX_DRUG_NOT_IN_LIBRARY or MDCX_DRUG_IS_IN_LIBRARY If not present, then consider it as 'not known'
- Update MDC_PUMP_PROGRAM_COMPLETION_MODE to add new enum for completion mode for switching sources: pump-program-completion-mode-switching-source or pump-program-completion-mode-switch-source
Decisions/Issues:
Action(s):
3 New
- tbdStatus/Discussion:
- MEMDMC containment
- add to agenda for Jul 14th, link copy of current file in agenda for prereview
- MDC_CLINICAL_USE_STATE Clinical use state could use more detail. i.e. Infusion pump is in use if it is programmed and ready to infuse. or the more generic case, device is in use if it is programmed/configured and ready to be used with a patient.
Decisions/Issues:
Action(s):
4 Next Meeting
- ChairStatus/Discussion: Decisions/Issues:
- Topics will include ...
Action(s):
- 3a. Multi-step/bolus order examples
Next Meeting
NOTE: Next Meeting will be held during the regularly scheduled time: 12:00 (Eastern) on July 14 (60 minutes)
(Reviewed & approved by PCD Infusion Pump WG on <TBD>