PCD Pump 2020-06-08 WebEx: Difference between revisions

From IHE Wiki
Jump to navigation Jump to search
Kelliason (talk | contribs)
Kelliason (talk | contribs)
 
(2 intermediate revisions by the same user not shown)
Line 32: Line 32:


:* '''<add attachments links here>'''
:* '''<add attachments links here>'''
: IPEC tests https://drive.google.com/open?id=16ZrSnb_Y4ZxhlntG1mPaJDjZK3DSVgz3
: Working copy of IPEC tests https://drive.google.com/file/d/1_PVitRThfnAxms0P7OhnfRrbEozjBu3n/view?usp=sharing


== Discussion Notes ==
== Discussion Notes ==
Line 38: Line 38:
:'''Chair/Host:'''  ''Kurt Elliason (Smiths Medical)''
:'''Chair/Host:'''  ''Kurt Elliason (Smiths Medical)''


:'''Participants:'''  '' tbd ''
:'''Participants:'''  '' Al Engelbert (BBraun), Bill Haralson (Smiths), Christophe Fournier (Fresenius), Jeff Rinda (ICU Medical), Marufur Rahmen (ICU), Paul Schluter (CMI), Sarah Johnston (BBraun), Tom Kowalczyk (BBraun)''


::'''NOTE:''' '''''Action Item Review updates have been posted directly to the [[PCD_Device_Infusion_Pump_Action_Items | Infusion Pump Action Items wiki page]]'''''
::'''NOTE:''' '''''Action Item Review updates have been posted directly to the [[PCD_Device_Infusion_Pump_Action_Items | Infusion Pump Action Items wiki page]]'''''
Line 61: Line 61:
|-
|-
| align="center" | 2
| align="center" | 2
| '''tbd''' <br>- tbd
| '''PCA Door Events''' <br>- Kurt
| '''Status/Discussion:'''
| '''Status/Discussion:'''




'''Decisions/Issues:'''
'''Decisions/Issues:'''
MDC_EVT_DROP_SENSOR_DISCONNECTION_REQD  - drops of liquid.  Sensor needs to be disconnected due to incompatibility with the selected drug.  Needs more consideration.


MDC_EVT_PCA_INSUF_DOSE should be MDC_EVT_TCI_INSUF_DOSE
Decided to use
MDC_EVT_TCI _VOL_LOW - In TCI mode, the remaining volume/dose in the syringe is insufficient to reach the target
- Not enough medication in the syringe to achieve the desired target.
Change MDC_EVT_PCA_MAX_BOLUSES  to MDC_EVT_PCA_MAX_BOLUS_COUNT_EXCEEDED
 
Having the PCA_NEAR_MAX_LIMIT companion to PCA_MAX_LIMIT aligns other term sets where we have both and “ending” and near “ending” events.


'''Action(s):'''
'''Action(s):'''
Line 73: Line 82:
|-
|-
| align="center" | 3
| align="center" | 3
| '''tbd''' <br>- tbd
| '''IPEC Test updates''' <br>- Kurt
| '''Status/Discussion:'''
| '''Status/Discussion:'''




'''Decisions/Issues:'''
'''Decisions/Issues:'''
 
Final version ready for PCD  review during Connectathon test review meeting.
https://drive.google.com/file/d/13QapYk0Air5WYsWMr65k-Hvun9770Mlg/view?usp=sharing


'''Action(s):'''
'''Action(s):'''

Latest revision as of 12:43, 8 June 2020

(Infusion Pump Main Page)


Meeting Purpose

IHE PCD Infusion Pump Working Group discussions.

WebEx Information

Topic: IHE PCD Infusion Pump WG

Date: Monday, June 8, 2020

Time: 11:00, Eastern Time (GMT -05:00, New York)

Duration: 60 Minutes


Note: Specific web & phone information will be provided via e-mail to group members.

Contact Kurt Elliason for more information.

Proposed Agenda

1. Review today's Agenda
2. Review 2020-05-11 Pump WebEx Session
3. Review PCA Door event terms
4. Review IPEC tests
5. Review PCD Device Infusion Pump Action Items
6. New Business / Next Meetings

Attachments / Materials

  • <add attachments links here>
Working copy of IPEC tests https://drive.google.com/file/d/1_PVitRThfnAxms0P7OhnfRrbEozjBu3n/view?usp=sharing

Discussion Notes

Chair/Host: Kurt Elliason (Smiths Medical)
Participants: Al Engelbert (BBraun), Bill Haralson (Smiths), Christophe Fournier (Fresenius), Jeff Rinda (ICU Medical), Marufur Rahmen (ICU), Paul Schluter (CMI), Sarah Johnston (BBraun), Tom Kowalczyk (BBraun)
NOTE: Action Item Review updates have been posted directly to the Infusion Pump Action Items wiki page


Item Topic Discussion
1 Agenda & Discussion Notes Review
- Chair
Status/Discussion:


Decisions/Issues:


Action(s):

2 PCA Door Events
- Kurt
Status/Discussion:


Decisions/Issues: MDC_EVT_DROP_SENSOR_DISCONNECTION_REQD - drops of liquid. Sensor needs to be disconnected due to incompatibility with the selected drug. Needs more consideration.

MDC_EVT_PCA_INSUF_DOSE should be MDC_EVT_TCI_INSUF_DOSE Decided to use MDC_EVT_TCI _VOL_LOW - In TCI mode, the remaining volume/dose in the syringe is insufficient to reach the target - Not enough medication in the syringe to achieve the desired target.

Change MDC_EVT_PCA_MAX_BOLUSES to MDC_EVT_PCA_MAX_BOLUS_COUNT_EXCEEDED   Having the PCA_NEAR_MAX_LIMIT companion to PCA_MAX_LIMIT aligns other term sets where we have both and “ending” and near “ending” events.

Action(s):


3 IPEC Test updates
- Kurt
Status/Discussion:


Decisions/Issues: Final version ready for PCD review during Connectathon test review meeting. https://drive.google.com/file/d/13QapYk0Air5WYsWMr65k-Hvun9770Mlg/view?usp=sharing

Action(s):


4 Next Meeting
- Chair
Status/Discussion:

Decisions/Issues:

  • Topics will include ...

Action(s):


Next Meeting

NOTE: Next Meeting will be held during the regularly scheduled time: 11:00 (Eastern) on June 22 (60 minutes)


(Reviewed & approved by PCD Infusion Pump WG on <TBD>


PCD Home