+ | <tr> | ||
+ | <td align="right" width="58"><div align="left">9001</div></td> | ||
+ | <td width="349">Unknown infuser or channel</td> | ||
+ | <td width="306">e.g. incorrect infuser or channel ID</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <td align="right" width="58"><div align="left">9002</div></td> | ||
+ | <td width="349">Infuser/channel is currently infusing</td> | ||
+ | <td width="306"></td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <td align="right" width="58"><div align="left">9003</div></td> | ||
+ | <td width="349">Missing required program parameter(s) (ParameterName1, ParameterName2, ...)</td> | ||
+ | <td width="306">e.g. Give Amount Minimum (RXG-5) - volume to be infused - is missing</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <td align="right" width="58"><div align="left">9004</div></td> | ||
+ | <td width="349">Invalid program parameter(s) (ParameterName1, ParameterName2, ...)</td> | ||
+ | <td width="306">e.g. volume units are not mL</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <td align="right" width="58"><div align="left">9005</div></td> | ||
+ | <td width="349">Parameter (ParameterName) outside of allowable range (MinValue to MaxValue)</td> | ||
+ | <td width="306">e.g. ordered rate greater than pump maximum</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <td align="right" width="58"><div align="left">9006</div></td> | ||
+ | <td width="349">Infuser is powered off</td> | ||
+ | <td width="306"></td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <td align="right" width="58"><div align="left">9007</div></td> | ||
+ | <td width="349">Infuser is offline or unable to connect to infuser</td> | ||
+ | <td width="306">e.g. infuser not on network or weak wireless signal</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <td align="right" width="58"><div align="left">9008</div></td> | ||
+ | <td width="349">Invalid units for parameter (ParameterName)</td> | ||
+ | <td width="306">e.g. Give Strength Volume Units (RXG-24) contains a medication unit value instead of volume units</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <td align="right" width="58"><div align="left">9009</div></td> | ||
+ | <td width="349">Dose/Rate Units do not match drug library</td> | ||
+ | <td width="306">e.g. ordered units = mL/hr; drug library units =mcg/kg/min</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <td align="right" width="58"><div align="left">9010</div></td> | ||
+ | <td width="349">Unable to match medication to drug library</td> | ||
+ | <td width="306">e.g. Medication does not exist in drug library</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <td align="right" width="58"><div align="left">9011</div></td> | ||
+ | <td width="349">Patient weight mismatch</td> | ||
+ | <td width="306">e.g. patient weight known by pump differs from weight sent in PCD-03</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <td align="right" width="58"><div align="left">9012</div></td> | ||
+ | <td width="349">Patient ID mismatch</td> | ||
+ | <td width="306">e.g. patient ID known by pump differs from ID sent in PCD-03</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <td align="right" width="58"><div align="left">9013</div></td> | ||
+ | <td width="349">Unable to program medication as piggyback</td> | ||
+ | <td width="306">e.g. medication not configured for piggyback administration in drug library</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <td align="right" width="58"><div align="left">9014</div></td> | ||
+ | <td width="349">Dose rate or VTBI exceeds maximum</td> | ||
+ | <td width="306">e.g. greater than pump maximum</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <td align="right" width="58"><div align="left">9015</div></td> | ||
+ | <td width="349">Request timed out</td> | ||
+ | <td width="306"></td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <td align="right" width="58"><div align="left">9016</div></td> | ||
+ | <td width="349">Unknown error</td> | ||
+ | <td width="306">Used when other errors are not applicable</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <td align="right" width="58"><div align="left">9017</div></td> | ||
+ | <td width="349">Infuser cannot accept program</td> | ||
+ | <td width="306">Infuser is in a state where it cannot accept program; e.g. alarming or in standby</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <td align="right" width="58"><div align="left">9018</div></td> | ||
+ | <td width="349">Parameter (ParameterName) does not match drug library </td> | ||
+ | <td width="306">e.g. Give Strength Units (RXG-18) = mg, drug library = mEq</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <td align="right" width="58"><div align="left">9019</div></td> | ||
+ | <td width="349">Patient weight missing</td> | ||
+ | <td width="306">Drug is weight-based or BSA-based, but patient weight OBX is missing</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <td align="right" width="58"><div align="left">9020</div></td> | ||
+ | <td width="349">Patient height missing</td> | ||
+ | <td width="306">Drug is BSA-based, but patient height OBX is missing </td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <td align="right" width="58"><div align="left">9021</div></td> | ||
+ | <td width="349">Care area or profile mismatch</td> | ||
+ | <td width="306">Care area or profile not cleared on pump; or pump is set to a different care area</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <td align="right" width="58"><div align="left">9022</div></td> | ||
+ | <td width="349">Requested infusion program is stale</td> | ||
+ | <td width="306">The value of ORC.9 is older than what the IOC will allow to program the pump</td> | ||
+ | </tr> | ||
+ | |||
+ | </table> |
Revision as of 14:43, 18 December 2012
Infusion Pump Error Codes (for PCD-03 Ack messages)
Underlying data files in various formats (.xml, .csv, Word) are in ftp://ftp.ihe.net/Patient_Care_Devices/Tools/PumpErrorCodes
Most recent update: 2012-12-18
9001 |
Unknown infuser or channel | e.g. incorrect infuser or channel ID |
9002 |
Infuser/channel is currently infusing | |
9003 |
Missing required program parameter(s) (ParameterName1, ParameterName2, ...) | e.g. Give Amount Minimum (RXG-5) - volume to be infused - is missing |
9004 |
Invalid program parameter(s) (ParameterName1, ParameterName2, ...) | e.g. volume units are not mL |
9005 |
Parameter (ParameterName) outside of allowable range (MinValue to MaxValue) | e.g. ordered rate greater than pump maximum |
9006 |
Infuser is powered off | |
9007 |
Infuser is offline or unable to connect to infuser | e.g. infuser not on network or weak wireless signal |
9008 |
Invalid units for parameter (ParameterName) | e.g. Give Strength Volume Units (RXG-24) contains a medication unit value instead of volume units |
9009 |
Dose/Rate Units do not match drug library | e.g. ordered units = mL/hr; drug library units =mcg/kg/min |
9010 |
Unable to match medication to drug library | e.g. Medication does not exist in drug library |
9011 |
Patient weight mismatch | e.g. patient weight known by pump differs from weight sent in PCD-03 |
9012 |
Patient ID mismatch | e.g. patient ID known by pump differs from ID sent in PCD-03 |
9013 |
Unable to program medication as piggyback | e.g. medication not configured for piggyback administration in drug library |
9014 |
Dose rate or VTBI exceeds maximum | e.g. greater than pump maximum |
9015 |
Request timed out | |
9016 |
Unknown error | Used when other errors are not applicable |
9017 |
Infuser cannot accept program | Infuser is in a state where it cannot accept program; e.g. alarming or in standby |
9018 |
Parameter (ParameterName) does not match drug library | e.g. Give Strength Units (RXG-18) = mg, drug library = mEq |
9019 |
Patient weight missing | Drug is weight-based or BSA-based, but patient weight OBX is missing |
9020 |
Patient height missing | Drug is BSA-based, but patient height OBX is missing |
9021 |
Care area or profile mismatch | Care area or profile not cleared on pump; or pump is set to a different care area |
9022 |
Requested infusion program is stale | The value of ORC.9 is older than what the IOC will allow to program the pump |