PCD Pump 2021-01-11 WebEx
(Redirected from PCD Pump 2020-01-11 WebEx)
Jump to navigation
Jump to search
Meeting Purpose
IHE PCD Infusion Pump Working Group discussions.
WebEx Information
Topic: IHE PCD Infusion Pump WG
Date: Monday, January 11, 2021
Time: 12:00, Eastern Time (GMT -05:00, New York)
Duration: 60 Minutes
Note: Specific web & phone information will be provided via e-mail to group members.
Contact Kurt Elliason for more information.
Proposed Agenda
- 1. Review today's Agenda
- 2. Review 2020-12-14 Pump WebEx Session
- 3. Discussion of FHIR based approach to pump programming - Brian W
- 4. CPs for new Rack terms
- CP for position in Pillar/Tower-Rack-Slot
- CP to document which terms are under the MDS if it is a Rack and which terms are under the pump's VMD.
- 5. Review PCD Device Infusion Pump Action Items
- 6. New Business / Next Meetings
Attachments / Materials
- <add attachments links here>
Discussion Notes
- Chair/Host: Kurt Elliason (Smiths Medical)
- Participants: Al Engelbert (BBraun), Andrea Robertson (Smiths), Brian Witkowski (BBraun), Jeff Rinda (Interoperability Solutions), Marufur Rahmen (ICU), Paul Schluter (CMI), Shantanu Chawla (BD), Tom Kowalczyk (BBraun)
- NOTE: Action Item Review updates have been posted directly to the Infusion Pump Action Items wiki page
Item Topic Discussion 1 Agenda & Discussion Notes Review
- ChairStatus/Discussion:
Decisions/Issues:
Action(s):2 Discussion of FHIR based approach to pump programming
- Brian W.Status/Discussion: - Deferred to future meeting
Decisions/Issues:
Action(s):
3 Rack CPs
- Kurt E.Status/Discussion: - CP for position in Pillar/Tower-Rack-Slot
- BBRaun reports Pillar number and slots in pillar
- Discussed option to create specific OBX’s for each. Eg.
- MDC_ATTR_RACK_INDEX – String - Left, Right, Top, Bottom
- MDC_ATTR_RACK_POSITION – String - Left, Right, Top, Bottom, Center -- RACK MDS
- MDC_ATTR_RACK_NUMBER – String - 1…X -- RACK MDS
- MDC_ATTR_RACK_CAPACITY – numeric - 1..X -- RACK MDS
- MDC_RACK_DETAILS_TERM – array - (index, number, capacity) i.e. (Left, 2, NOS)
- MDC_ATTR_PUMP_RACK_POSITION – numeric - 1..X PUMP VMD
- Or
- MDC_PUMPRACK_DETAILS_TERM – array - (index, number, capacity, slot) i.e. (Left, 2, NOS, 3)
- Could have Two REFIDS – maximal position, actual position
Decisions/Issues:- Picture for different options -Tom/Al will diagram how theirs works, others can provide thiers prior to next meeting for discussion.
- 1 vs 1-2-3-4
- 2
- 3
- 4
Action(s):
- CP to document which terms are under the MDS if it is a Rack and which terms are under the pump's VMD. Provide a picture.
4 Next Meeting
- ChairStatus/Discussion: Decisions/Issues:
- Topics will include ...
Action(s):
Next Meeting
NOTE: Next Meeting will be held during the regularly scheduled time: 11:00 (Eastern) on January 25 (60 minutes)
(Reviewed & approved by PCD Infusion Pump WG on <TBD>