Overview

The IHE PCD Technical Framework is composed of a set of profiles that address specific integration scenarios that include at least one regulated medical device (per the PCD Charter).

<volume 1> <volume 2> <volume 3>

Current Framework Status

First published in the summer of 2006, the current PCD technical framework ...

PCD Transaction Map

Many of the PCD Profiles include transactions, specified in the Technical Framework Volume 2 Transactions, which are allocated on a domain basis. This table identifies the currently allocated transactions and the profiles to which they belong:


ID Transaction Title Profile Source Actor Receiving Actor Notes
PCD-01 Communicate PCD Data DEC Device Observation Reporter (DOR) Device Observation Consumer (DOC)
PCD-02 Subscribe to PCD Data SPD Device Observation Consumer (DOC) Device Observation Filter (DOF)
PCD-03 Communicate Infusion Order PIV Infusion Order Programmer (IOP) Infusion Order Consumer (IOC)
PCD-04 Report Alarm ACM Alarm Reporter (AR) Alarm Manager (AM) Formerly ACM-01
PCD-05 Report Alarm Status ACM Alarm Manager (AM) Alarm Reporter (AR) Formerly ACM-02
PCD-06 Disseminate Alarm ACM Alarm Manager (AM) Alarm Communicator (AC) Formerly ACM-03
PCD-07 Report Dissemination Alarm Status ACM Alarm Communicator (AC) Alarm Manager (AM) Formerly ACM-04