PaLM Conf Minutes 2026-July-08: Difference between revisions
Riki Merrick (talk | contribs) No edit summary |
Riki Merrick (talk | contribs) No edit summary |
||
| (One intermediate revision by the same user not shown) | |||
| Line 18: | Line 18: | ||
*UDI in LAW and LTW | *UDI in LAW and LTW | ||
[Changed UDI Discussion PaLM F2F 2026-06-16 (1).pptx|UDI in HL7] | [[:File:Changed UDI Discussion PaLM F2F 2026-06-16 (1).pptx|UDI in HL7 presentation]] | ||
**The production identifiers used by Abbott for reagents matches what Roche is doing – does it also apply to BioMérieux? | **The production identifiers used by Abbott for reagents matches what Roche is doing – does it also apply to BioMérieux? | ||
**In LAW we send manufacturer and model name – with this information you could look up the device ID in LIVD or GUUID, though we could make a CR to include the GTIN in the first repeat of OBX-18.3 for the DI, OR could make update and remove required repeat OBX-18 for the PI and send the full UDI in OBX-18.3 | **In LAW we send manufacturer and model name – with this information you could look up the device ID in LIVD or GUUID, though we could make a CR to include the GTIN in the first repeat of OBX-18.3 for the DI, OR could make update and remove required repeat OBX-18 for the PI and send the full UDI in OBX-18.3 | ||
Latest revision as of 06:56, 21 October 2026
Attendees:
Ralf Herzog (Roche), Alessandro Sulis, Kevin Schap (CAP), Ruben Fernandes (Sectra), Gunter Haroske, Corinne (Biomerieux), David Clunie, Philippe Fournier (Biomerieux), James Howe, Megumi Kondo, Patty Hernandez, Kazunari Fukagawa, Robert Rae (CAP), James Harrison (CAP), Dan Rutz (Epic), Rajesh Bussa, Raj Dash (Duke/CAP), Ed Heierman (Abbott), Riki Merrick
Agenda:
• Last F2F wrap-up/comments/actions • DPIA: CPs ballot and planning for next version (1.5 ) • PaLM profile candidates for EU CAT 2027 • Projects updates • AOB • UDI in LAW and LTW
Notes:
- LAW CP for AutoDilution:
- Discussion as to why not convert 2/3 to a decimal – level of precision (though with 13 decimal points, probably precise enough) but also this is often how solutions are combined, so better matches the documentation in the lab
- Discussion about length – for first number 15 is too much, maybe 5 – decided to use 2 in the end
- Make length 2 and add comment that the first number should be a positive number, often ‘1’
- Will include the ballot that just went out by EOD today, so that voting can still close on August 7
- UDI in LAW and LTW
- The production identifiers used by Abbott for reagents matches what Roche is doing – does it also apply to BioMérieux?
- In LAW we send manufacturer and model name – with this information you could look up the device ID in LIVD or GUUID, though we could make a CR to include the GTIN in the first repeat of OBX-18.3 for the DI, OR could make update and remove required repeat OBX-18 for the PI and send the full UDI in OBX-18.3
- Next Steps:
- Bring to IICC to ask which option they prefer
- Write CP for LAW
- For LTW should use PRT for both instrument and reagents and everything else jurisdictions might want to include
- Next steps:
- Write CP for LTW (Jim is writing the message structure for the PMW profile and we could use that as the profile to be added as option to LTW)
- Ralf and Riki will pick this up early August
- Will then send for review to Ed, Xavier, Dan and check with LIDR group
- Reach out to more LIS vendors: SysteLab, Clinisys, Oracle, Riki’s contacts
- Create list of codes to be added to https://terminology.hl7.org/en/CodeSystem-v2-0912.html as siblings to EQUP (which is reserved for OBX-18 entries) to represent: instrument, reagent, calibrator and any others
- Last F2F wrap-up/comments/actions
- DPOW status still as at the end of F2F – needs to get migrated to IHE Supplement template
- Had a relatively quick meeting between IHE PaLM to DICOM WG26
- ADD RAJ’s SLIDE!
- Indicating if electronic or physical – using OBR-4 values to identify the workorder step types
- Automation of QC, so need DICOM elements to cover those (Barco’s presentation); could use SPM-21 in V2
- Maybe we need a DPQC profile (split this out from DPEC) – would need to get Barco to participate; need to consider when the pathologist needs a new slide (recut), technologist might explain why there won’t be anything better
- Could we use SDC templates for evidence creation output for each type of algorithm
- Can DICOM SR support the hierarchy within the SDC encapsulation, without having to recreate the hierarchy in DICOM; currently the evidence creators are looking into using FHIR Observation
- DICOM WG 26 Connectathon Update:
- For NA IHE Connectathon registration is still open till July 15, so far only have 1 vendor signed up for DPIA
- Have good number of scanners and image displays
- 2 acquisition managers
- Have a few evidence creators
- Focus on
- z-stacks and ???<missed what David said here>
- New compression schemes
- HL7 transactions are focusing on how to encode the stains and in the process found several that are missing in SNOMED CT, Hung Luu is now the liaison to SCT, they have agreed on a model on how to add those codes,
- SCT code use is now free to use (not the model behind it), so this is great news 😊
- DPIA: CPs ballot and planning for next version (1.5 )
- These were just now sent out to the IHE PaLM list
- Please vote – due by August 7, 2026
From Chat:
Raj Dash, MD (Duke Pathology) 9:11 AM Sorry to be late!
Rúben Fernandes (Sectra) 9:09 AM What's the smallest prime number with 5 digits? :)
Kevin Schap 9:13 AM 10007
Raj Dash, MD (Duke Pathology) 9:29 AM Might as well pre-adopt PRT and offer in guidance. As long as everyone does it the same way is all that matters... Rúben Fernandes (Sectra) 9:38 AM And Roy has moved to Barco Dedalus? Alessandro Sulis 9:44 AM https://drive.google.com/drive/u/0/folders/1q-p_efI8foXstZbm0t0NIsUuJoJgVdWZ @Ruben this is the Templates directory, from the IHE google drive
Rúben Fernandes (Sectra) 9:45 AM Thanks Alessandro
David Clunie 9:49 AM Diricx, Bart bart.diricx@barco.com
Kevin Schap 9:51 AM email sent, included David and Ruben
Raj Dash, MD (Duke Pathology) 9:54 AM I have a hard stop at the top of the hour James Harrison 9:58 AM I also need to drop off for a 10 am EDT meeting. See you next month and I look forward to hearing from the group looking at PRT in LTW> Rúben Fernandes (Sectra) 10:00 AM Maybe resuscitate this: JSON Representation of DICOM Structured Reports
Patty Hernandez (she/her) 10:00 AM I need to drop off for another meeting. Thanks!