PCD Pump 2025-07-14 Teams: Difference between revisions
Jump to navigation
Jump to search
| (One intermediate revision by the same user not shown) | |||
| Line 24: | Line 24: | ||
:1. Review today's Agenda | :1. Review today's Agenda | ||
:2. Review [[PCD_Pump_2025-06-28_Teams | 2025-06-28 Pump Teams Session]] | :2. Review [[PCD_Pump_2025-06-28_Teams | 2025-06-28 Pump Teams Session]] | ||
:3a | :3a. Rack/Stack [https://docs.google.com/document/d/1NQKnAvu9d9MfiTO0929O6NCfm7DaDcUs/edit?usp=sharing&ouid=108127959796856836161&rtpof=true&sd=true CP-DEV-020 draft] | ||
:3b. New terms - drug not in library and completion mode for switching sources | |||
: | :3c. [https://drive.google.com/file/d/1H8idlyAoEwkPuemLLvLIh_TC0ZJaAhkI/view?usp=sharing pcd-15 containment for pumps] | ||
:4a. PCD-03 Error codes | :4a. PCD-03 Error codes | ||
:4b. White paper updates [https://docs.google.com/document/d/1kbFbmYOjYphtXxW6MySFcfTviy2qwYNS/edit?usp=sharing&ouid=108127959796856836161&rtpof=true&sd=true Device Identification June draft] | :4b. White paper updates [https://docs.google.com/document/d/1kbFbmYOjYphtXxW6MySFcfTviy2qwYNS/edit?usp=sharing&ouid=108127959796856836161&rtpof=true&sd=true Device Identification June draft] | ||
:4c | :4c. PCD-03 Clinician Bolus (ORC-1 = CH) | ||
:5. Review [[PCD Device Infusion Pump Action Items | PCD Device Infusion Pump Action Items]] | :5. Review [[PCD Device Infusion Pump Action Items | PCD Device Infusion Pump Action Items]] | ||
:6. New Business / Next Meetings | :6. New Business / Next Meetings | ||
| Line 45: | Line 44: | ||
:'''Chair/Host:''' ''Kurt Elliason (ICU Medical)'' | :'''Chair/Host:''' ''Kurt Elliason (ICU Medical)'' | ||
:'''Participants:''' '' | :'''Participants:''' ''Al Engelbert (BBraun), Amy Rowe (BBraun), Brian Witkowski (Baxter), Evan O'Grady(Fresenius-Kabi), Michael Workman (BD), Nicolas Crouzier (NIST), Paul Schluter (Paul Schluter Consulting), Shantanu Chawla (BD), Silas Morris (Epic), Tom Kowalczyk (BBraun) | ||
| Line 72: | Line 72: | ||
| ''' Agenda topics''' <br>- tbd | | ''' Agenda topics''' <br>- tbd | ||
| '''Status/Discussion:''' | | '''Status/Discussion:''' | ||
:3a. Rack/Stack | |||
:: CP-DEV-020 - not ready for ballot, checked with Michael F. & Nicolas C. about NIST Connectathon tool error when using 2^^7. | |||
:::OBX-5 is a repeatable element so ~ are allowed here. | |||
:::And since you are using ST, a primitive, ^ will not make sense and are not allowed. | |||
::Should we use a different data type instead of ST. or use ST and not use empty strings with ~ (2,NA,NA,7) is 2~NA~NA~7 instead of 2^^^7 | |||
:3b. New terms | |||
::Proposed new term for PCD-01/-10: MDCX_DRUG_IS_IN_LIBRARY | |||
::: · Boolean value indicating if the currently programmed drug is in the drug library | |||
::: · If not present, assumed to be unknown | |||
::: · Rationale: Manually programmed drugs will always be in the drug library, but auto-programmed drugs may not be if they are experimental or the drug library has not been updated to include the new drug(s). It is important to know if drugs outside of the library are being administered, since those drugs will not have limits enforced. In addition, reporting if the drug is in the drug library will allow the EMR to monitor alignment between their formulary and the drug library. | |||
· | |||
::Proposed new value for MDC_PUMP_PROGRAM_COMPLETION_MODE: pump-program-completion-mode-switch-source | |||
::: · Description/Definition of MDC_PUMP_PROGRAM_COMPLETION_MODE: Programmed operation of the pump when a delivery completes (e.g., KVO). | |||
::: · Currently defined set of values: | |||
:::: o pump-program-completion-mode-stop | |||
:::: o pump-program-completion-mode-kvo | |||
:::: o pump-program-completion-mode-continue | |||
::: · None of the currently defined values address the case where a pump that is running a secondary will automatically switch to the primary when the secondary is complete | |||
::: · Note: the set of values for MDC_PUMP_NOT_DELIVERING_REASON includes pump-stopped-switching-source | |||
:3c. PCD-15 containment for pumps | |||
:: MDC_CLINICAL_USE_STATE Clinical use state could use more detail. i.e. Infusion pump is in use if it is programmed and ready to infuse. or the more generic case, device is in use if it is programmed/configured and ready to be used with a patient. | |||
:4a. PCD-03 Error codes | |||
:: No updates | |||
:4b. White paper updates | |||
:: Ready to review with DEV Technical Committee | |||
:4c. PCD-03 Clinician Bolus (ORC-1 = CH) | |||
::Discussed possible option of having multiple RXG segments like a multistep program for Loading Dose, Continuous, PCA etc but without steps. Add to agenda to review next meeting. | |||
Latest revision as of 18:11, 17 July 2025
Meeting Purpose
IHE PCD Infusion Pump Working Group discussions.
WebEx Information
Topic: IHE PCD Infusion Pump WG
Date: Monday, July 14, 2025
Time: 12:00, Eastern Time (GMT -05:00, New York)
Duration: 60 Minutes
Note: Specific web & phone information will be provided via e-mail to group members.
Contact Kurt Elliason for more information.
Proposed Agenda
- 1. Review today's Agenda
- 2. Review 2025-06-28 Pump Teams Session
- 3a. Rack/Stack CP-DEV-020 draft
- 3b. New terms - drug not in library and completion mode for switching sources
- 3c. pcd-15 containment for pumps
- 4a. PCD-03 Error codes
- 4b. White paper updates Device Identification June draft
- 4c. PCD-03 Clinician Bolus (ORC-1 = CH)
- 5. Review PCD Device Infusion Pump Action Items
- 6. New Business / Next Meetings
Attachments / Materials
- <add attachments links here>
Discussion Notes
- Chair/Host: Kurt Elliason (ICU Medical)
- Participants: Al Engelbert (BBraun), Amy Rowe (BBraun), Brian Witkowski (Baxter), Evan O'Grady(Fresenius-Kabi), Michael Workman (BD), Nicolas Crouzier (NIST), Paul Schluter (Paul Schluter Consulting), Shantanu Chawla (BD), Silas Morris (Epic), Tom Kowalczyk (BBraun)
- NOTE: Action Item Review updates have been posted directly to the Infusion Pump Action Items wiki page
Item Topic Discussion 1 Agenda & Discussion Notes Review
- ChairStatus/Discussion:
Decisions/Issues:
Action(s):2 Agenda topics
- tbdStatus/Discussion: - 3a. Rack/Stack
- CP-DEV-020 - not ready for ballot, checked with Michael F. & Nicolas C. about NIST Connectathon tool error when using 2^^7.
- OBX-5 is a repeatable element so ~ are allowed here.
- And since you are using ST, a primitive, ^ will not make sense and are not allowed.
- Should we use a different data type instead of ST. or use ST and not use empty strings with ~ (2,NA,NA,7) is 2~NA~NA~7 instead of 2^^^7
- CP-DEV-020 - not ready for ballot, checked with Michael F. & Nicolas C. about NIST Connectathon tool error when using 2^^7.
- 3b. New terms
- Proposed new term for PCD-01/-10: MDCX_DRUG_IS_IN_LIBRARY
- · Boolean value indicating if the currently programmed drug is in the drug library
- · If not present, assumed to be unknown
- · Rationale: Manually programmed drugs will always be in the drug library, but auto-programmed drugs may not be if they are experimental or the drug library has not been updated to include the new drug(s). It is important to know if drugs outside of the library are being administered, since those drugs will not have limits enforced. In addition, reporting if the drug is in the drug library will allow the EMR to monitor alignment between their formulary and the drug library.
- Proposed new term for PCD-01/-10: MDCX_DRUG_IS_IN_LIBRARY
·
- Proposed new value for MDC_PUMP_PROGRAM_COMPLETION_MODE: pump-program-completion-mode-switch-source
- · Description/Definition of MDC_PUMP_PROGRAM_COMPLETION_MODE: Programmed operation of the pump when a delivery completes (e.g., KVO).
- · Currently defined set of values:
- o pump-program-completion-mode-stop
- o pump-program-completion-mode-kvo
- o pump-program-completion-mode-continue
- · None of the currently defined values address the case where a pump that is running a secondary will automatically switch to the primary when the secondary is complete
- · Note: the set of values for MDC_PUMP_NOT_DELIVERING_REASON includes pump-stopped-switching-source
- Proposed new value for MDC_PUMP_PROGRAM_COMPLETION_MODE: pump-program-completion-mode-switch-source
- 3c. PCD-15 containment for pumps
- MDC_CLINICAL_USE_STATE Clinical use state could use more detail. i.e. Infusion pump is in use if it is programmed and ready to infuse. or the more generic case, device is in use if it is programmed/configured and ready to be used with a patient.
- 4a. PCD-03 Error codes
- No updates
- 4b. White paper updates
- Ready to review with DEV Technical Committee
- 4c. PCD-03 Clinician Bolus (ORC-1 = CH)
- Discussed possible option of having multiple RXG segments like a multistep program for Loading Dose, Continuous, PCA etc but without steps. Add to agenda to review next meeting.
Decisions/Issues:
Action(s):
3 TBD
- tbdStatus/Discussion:
Decisions/Issues:
Action(s):
4 Next Meeting
- ChairStatus/Discussion: Decisions/Issues:
- Topics will include ...
Action(s):
- 3a. Rack/Stack
Next Meeting
NOTE: Next Meeting will be held during the regularly scheduled time: 12:00 (Eastern) on July 28 (60 minutes)
(Reviewed & approved by PCD Infusion Pump WG on <TBD>