− | + | Radiotherapy Treatment WorkFlow (TRWF) integrates daily imaging with radiation therapy treatments using workflow. | |
− | + | ||
__TOC__ | __TOC__ | ||
Line 6: | Line 6: | ||
==Summary== | ==Summary== | ||
Two Workflows are described by this profile: | Two Workflows are described by this profile: | ||
− | * | + | * Discrete Positioning and Delivery Workflow involves positioning a patient and subsequent treatment delivery performed by separate devices. |
− | * | + | * Integrated Positioning and Delivery Workflow involves positioning a patient and subsequent treatment delivery, both performed by a single device. |
==Benefits== | ==Benefits== | ||
− | * | + | * interoperability between multi-vendor radiotherapy systems |
− | |||
− | |||
==Details== | ==Details== | ||
Line 29: | Line 27: | ||
* Treatment Planning Systems | * Treatment Planning Systems | ||
* Radiotherapy Treatment Delivery Systems | * Radiotherapy Treatment Delivery Systems | ||
+ | |||
+ | '''Actors & Transactions: Discrete Positioning and Delivery Workflow''' | ||
+ | [[image:TRWF discrete transactions.jpg]] | ||
+ | |||
+ | '''Actors & Transactions: Integrated Positioning and Delivery Workflow''' | ||
+ | [[image:TRWF integrated transactions.jpg]] | ||
==Specification== | ==Specification== |
Revision as of 18:17, 16 May 2008
Radiotherapy Treatment WorkFlow (TRWF) integrates daily imaging with radiation therapy treatments using workflow.
Summary
Two Workflows are described by this profile:
- Discrete Positioning and Delivery Workflow involves positioning a patient and subsequent treatment delivery performed by separate devices.
- Integrated Positioning and Delivery Workflow involves positioning a patient and subsequent treatment delivery, both performed by a single device.
Benefits
- interoperability between multi-vendor radiotherapy systems
Details
The Discrete Positioning and Delivery Workflow Profile involves the positioning of a patient and subsequent treatment delivery, where the positioning and delivery are performed by separate devices, as follows:
- The Patient Positioning System (PPS) acquires a set of 2D (planar projection) or 3D (CT) positioning images, performs a registration with previously retrieved reference images, and repositions the patient if necessary. Note that for this profile, the output of these steps is the fact that the patient positioning has been successfully performed.
- The Treatment Delivery Device (TDD) then delivers the intended treatment.
The Integrated Positioning and Delivery Workflow Profile involves the positioning of a patient and subsequent treatment delivery, where the positioning and delivery are performed by a single device, as follows:
- The Positioning and Delivery System (PDS) acquires a set of 2D (planar projection) or 3D (CT) positioning images, performs a registration with previously retrieved reference images, and repositions the patient if necessary. It then delivers the intended treatment.
Systems Affected
- Radiation Oncology PACS
- Treatment Planning Systems
- Radiotherapy Treatment Delivery Systems
Actors & Transactions: Discrete Positioning and Delivery Workflow
Actors & Transactions: Integrated Positioning and Delivery Workflow
Specification
Profile Status: Final Text
Documents:
IHE Radiation Oncology Technical Framework:
- Managed Delivery Workflow - Chapter 3 & 4 and Appendix A & B
- see DICOM Supplement 74
Underlying Standards:
See Also
Related Profiles
- Normal Treatment Planning-Simple (NTPL-S)
- Multimodality Registration for Radiation Oncology (MMR-RO)
Consumer Information
- The TRWF FAQ answers typical questions about what the Profile does.
- The TRWF Purchasing describes considerations when purchasing equipment to deploy this Profile.
Implementer Information
- The TRWF Implementation provides additional information about implementing this Profile in software.
This page is based on the Profile Template