Difference between revisions of "PumpErrorCodes"

From IHE Wiki
Jump to navigation Jump to search
(add codes 9018-9022)
m
Line 2: Line 2:
 
[[Category: PCD Reference Pages]]
 
[[Category: PCD Reference Pages]]
  
==Infusion Pump Error Codes (for PCD-03 Ack messages)==
+
==PIV Error Codes(for use in PCD-03 Ack messages)==
  
 
Underlying data files in various formats (.xml, .csv, Word) are in ftp://ftp.ihe.net/Patient_Care_Devices/Tools/PumpErrorCodes
 
Underlying data files in various formats (.xml, .csv, Word) are in ftp://ftp.ihe.net/Patient_Care_Devices/Tools/PumpErrorCodes

Revision as of 13:47, 12 November 2013


PIV Error Codes(for use in PCD-03 Ack messages)

Underlying data files in various formats (.xml, .csv, Word) are in ftp://ftp.ihe.net/Patient_Care_Devices/Tools/PumpErrorCodes

Most recent update: 2012-12-18

9001
Unknown infuser or channel e.g. incorrect infuser or channel ID
9002
Infuser/channel is currently infusing
9003
Missing required program parameter(s) (ParameterName1, ParameterName2, ...) e.g. Give Amount Minimum (RXG-5) - volume to be infused - is missing
9004
Invalid program parameter(s) (ParameterName1, ParameterName2, ...) e.g. volume units are not mL
9005
Parameter (ParameterName) outside of allowable range (MinValue to MaxValue) e.g. ordered rate greater than pump maximum
9006
Infuser is powered off
9007
Infuser is offline or unable to connect to infuser e.g. infuser not on network or weak wireless signal
9008
Invalid units for parameter (ParameterName) e.g. Give Strength Volume Units (RXG-24) contains a medication unit value instead of volume units
9009
Dose/Rate Units do not match drug library e.g. ordered units = mL/hr; drug library units =mcg/kg/min
9010
Unable to match medication to drug library e.g. Medication does not exist in drug library
9011
Patient weight mismatch e.g. patient weight known by pump differs from weight sent in PCD-03
9012
Patient ID mismatch e.g. patient ID known by pump differs from ID sent in PCD-03
9013
Unable to program medication as piggyback e.g. medication not configured for piggyback administration in drug library
9014
Dose rate or VTBI exceeds maximum e.g. greater than pump maximum
9015
Request timed out
9016
Unknown error Used when other errors are not applicable
9017
Infuser cannot accept program Infuser is in a state where it cannot accept program; e.g. alarming or in standby
9018
Parameter (ParameterName) does not match drug library    e.g. Give Strength Units (RXG-18) = mg, drug library = mEq
9019
Patient weight missing Drug is weight-based or BSA-based, but patient weight OBX is missing
9020
Patient height missing Drug is BSA-based, but patient height OBX is missing 
9021
Care area or profile mismatch Care area or profile not cleared on pump; or pump is set to a different care area
9022
Requested infusion program is stale The value of ORC.9 is older than what the IOC will allow to program the pump