Difference between revisions of "PumpErrorCodes"

From IHE Wiki
Jump to navigation Jump to search
(add heading, codes per CP 116)
 
Line 3: Line 3:
  
 
==PIV Error Codes (for use in PCD-03 Ack messages)==
 
==PIV Error Codes (for use in PCD-03 Ack messages)==
 
+
:See appendix B.3 ERR-Error Segment in "IHE Patient Care Device Technical Framework, Vol. 2 (PCD TF-2): Transactions" (https://www.ihe.net/uploadedFiles/Documents/PCD/IHE_PCD_TF_Vol2.pdf) for specific error codes.
Underlying data files in various formats (.xml, .csv, Word) are in ftp://ftp.ihe.net/Patient_Care_Devices/Tools/PumpErrorCodes
 
 
 
Most recent update: 2012-12-18
 
 
 
<table style="border-collapse:collapse; border-style:solid; border-width:1px; padding:0.5em; vertical-align:text-top; " border="1">
 
 
 
  <tr>
 
    <td align="right" width="58"><div align="left">Code</div></td>
 
    <td width="349">Text</td>
 
    <td width="306">Example</td>
 
  </tr>
 
 
 
  <tr>
 
    <td align="right" width="58"><div align="left">9001</div></td>
 
    <td width="349">Unknown infuser or channel</td>
 
    <td width="306">e.g. incorrect infuser or channel ID</td>
 
  </tr>
 
  <tr>
 
    <td align="right" width="58"><div align="left">9002</div></td>
 
    <td width="349">Infuser/channel is    currently infusing</td>
 
    <td width="306"></td>
 
  </tr>
 
  <tr>
 
    <td align="right" width="58"><div align="left">9003</div></td>
 
    <td width="349">Missing required program parameter(s)    (ParameterName1, ParameterName2, ...)</td>
 
    <td width="306">e.g. Give Amount Minimum (RXG-5) - volume to be    infused - is missing</td>
 
  </tr>
 
  <tr>
 
    <td align="right" width="58"><div align="left">9004</div></td>
 
    <td width="349">Invalid program    parameter(s) (ParameterName1, ParameterName2, ...)</td>
 
    <td width="306">e.g. volume units are not    mL</td>
 
  </tr>
 
  <tr>
 
    <td align="right" width="58"><div align="left">9005</div></td>
 
    <td width="349">Parameter (ParameterName) outside of allowable    range (MinValue to MaxValue)</td>
 
    <td width="306">e.g. ordered rate greater than pump maximum</td>
 
  </tr>
 
  <tr>
 
    <td align="right" width="58"><div align="left">9006</div></td>
 
    <td width="349">Infuser is powered off</td>
 
    <td width="306"></td>
 
  </tr>
 
  <tr>
 
    <td align="right" width="58"><div align="left">9007</div></td>
 
    <td width="349">Infuser is offline or unable to connect to infuser</td>
 
    <td width="306">e.g. infuser not on network or weak wireless signal</td>
 
  </tr>
 
  <tr>
 
    <td align="right" width="58"><div align="left">9008</div></td>
 
    <td width="349">Invalid units for    parameter (ParameterName)</td>
 
    <td width="306">e.g. Give Strength Volume    Units (RXG-24) contains a medication unit value instead of volume units</td>
 
  </tr>
 
  <tr>
 
    <td align="right" width="58"><div align="left">9009</div></td>
 
    <td width="349">Dose/Rate Units do not match drug library</td>
 
    <td width="306">e.g. ordered units = mL/hr; drug library units    =mcg/kg/min</td>
 
  </tr>
 
  <tr>
 
    <td align="right" width="58"><div align="left">9010</div></td>
 
    <td width="349">Unable to match medication    to drug library</td>
 
    <td width="306">e.g. Medication does not    exist in drug library</td>
 
  </tr>
 
  <tr>
 
    <td align="right" width="58"><div align="left">9011</div></td>
 
    <td width="349">Patient weight mismatch</td>
 
    <td width="306">e.g. patient weight known by pump differs from    weight sent in PCD-03</td>
 
  </tr>
 
  <tr>
 
    <td align="right" width="58"><div align="left">9012</div></td>
 
    <td width="349">Patient ID mismatch</td>
 
    <td width="306">e.g. patient ID known by    pump differs from ID sent in PCD-03</td>
 
  </tr>
 
  <tr>
 
    <td align="right" width="58"><div align="left">9013</div></td>
 
    <td width="349">Unable to program medication as piggyback</td>
 
    <td width="306">e.g. medication not configured for piggyback    administration in drug library</td>
 
  </tr>
 
  <tr>
 
    <td align="right" width="58"><div align="left">9014</div></td>
 
    <td width="349">Dose rate or VTBI exceeds    maximum</td>
 
    <td width="306">e.g. greater than pump    maximum</td>
 
  </tr>
 
  <tr>
 
    <td align="right" width="58"><div align="left">9015</div></td>
 
    <td width="349">Request timed out</td>
 
    <td width="306"></td>
 
  </tr>
 
  <tr>
 
    <td align="right" width="58"><div align="left">9016</div></td>
 
    <td width="349">Other error</td>
 
    <td width="306">Used when other errors are not applicable</td>
 
  </tr>
 
  <tr>
 
    <td align="right" width="58"><div align="left">9017</div></td>
 
    <td width="349">Infuser cannot accept program</td>
 
    <td width="306">Infuser is in a state where it cannot accept    program; e.g. alarming or in standby</td>
 
  </tr>
 
  <tr>
 
    <td align="right" width="58"><div align="left">9018</div></td>
 
    <td width="349">Parameter (ParameterName)    does not match drug library   </td>
 
    <td width="306">e.g. Give Strength Units    (RXG-18) = mg, drug library = mEq</td>
 
  </tr>
 
  <tr>
 
    <td align="right" width="58"><div align="left">9019</div></td>
 
    <td width="349">Patient weight missing</td>
 
    <td width="306">Drug is weight-based or BSA-based, but patient    weight OBX is missing</td>
 
  </tr>
 
  <tr>
 
    <td align="right" width="58"><div align="left">9020</div></td>
 
    <td width="349">Patient height missing</td>
 
    <td width="306">Drug is BSA-based, but    patient height OBX is missing </td>
 
  </tr>
 
  <tr>
 
    <td align="right" width="58"><div align="left">9021</div></td>
 
    <td width="349">Care area or profile mismatch</td>
 
    <td width="306">Care area or profile not cleared on pump; or pump    is set to a different care area</td>
 
  </tr>
 
  <tr>
 
    <td align="right" width="58"><div align="left">9022</div></td>
 
    <td width="349">Requested infusion program    is stale</td>
 
    <td width="306">The value of ORC.9 is    older than what the IOC will allow to program the pump</td>
 
  </tr>
 
  <tr>
 
    <td align="right" width="58"><div align="left">9023</div></td>
 
    <td width="349">Program rejected by user</td>
 
    <td width="306">Program rejected by user prior to starting infusion</td>
 
  </tr>
 
  <tr>
 
    <td align="right" width="58"><div align="left">9024</div></td>
 
    <td width="349">Drug library hard limit exceeded</td>
 
    <td width="306">e.g. dose exceeds maximum allowable for medication</td>
 
  </tr>
 
 
 
</table>
 

Latest revision as of 11:10, 9 July 2018


PIV Error Codes (for use in PCD-03 Ack messages)

See appendix B.3 ERR-Error Segment in "IHE Patient Care Device Technical Framework, Vol. 2 (PCD TF-2): Transactions" (https://www.ihe.net/uploadedFiles/Documents/PCD/IHE_PCD_TF_Vol2.pdf) for specific error codes.