PCD Pump 2022-01-10 Zoom
Jump to navigation
Jump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.
Meeting Purpose
IHE PCD Infusion Pump Working Group discussions.
WebEx Information
Topic: IHE PCD Infusion Pump WG
Date: Monday, Jan 10, 2022
Time: 12:00, Eastern Time (GMT -05:00, New York)
Duration: 60 Minutes
Note: Specific web & phone information will be provided via e-mail to group members.
Contact Kurt Elliason for more information.
Proposed Agenda
- 1. Review today's Agenda
- 2. Review 2021-12-13 Pump Zoom Session
- 3. Discussion of MDC_FLOW_FLUID_PUMP, MDC_FLOW_FLUID_PUMP_CURRENT, MDC_FLOW_FLUID_PUMP_SETTING
- 4. Review containment draft for pcd-01 - enumerations
- 5. Review PCD Device Infusion Pump Action Items
- 6. New Business / Next Meetings
Attachments / Materials
- <add attachments links here>
- PCD03 containment for discussion
Discussion Notes
- Chair/Host: Kurt Elliason (Smiths Medical)
- Participants: Adam Bunton(Cerner), Al Engelbert (BBraun), Andrea Robertson (Smiths), Brian Sullivan(BD), Brian Witkowski (BBraun), Jeff Rinda (Interoperability Solutions), Paul Schluter (), Rachel Golan (Root3 Labs), Steve Dain(), Tom Kowalczyk (BBraun);
- NOTE: Action Item Review updates have been posted directly to the Infusion Pump Action Items wiki page
Item Topic Discussion 1 Agenda & Discussion Notes Review
- ChairStatus/Discussion:
Decisions/Issues:
Action(s):2 MDC_FLOW_FLUID_PUMP
- Chair + Steve DainStatus/Discussion: - Existing term MDC_FLOW_FLUID_PUMP 26712 (secondary refid as MDC_RATE_INFUS) which is currently defined as "Volume rate delivered by pump" in the 10101:2019 and is defined differently as "Programmed volumetric delivery rate" in latest PCD10 containment spreadsheet
- New proposed term MDC_FLOW_FLUID_PUMP_CURRENT code 26942 defined as "current pump fluid flow"
- Historically we have used 26712 as the flow of the pump.
- Pump WG desired to communicate what the pump rate is set at, even when the pump is not infusing.
- We may have created some difficulties with the proposed term and the "new" definition of an old term.
- Suggestion is to delete the proposed term MDC_FLOW_FLUID_PUMP_CURRENT, clarifying the definition of MDC_FLOW_FLUID_PUMP 26712 as "current fluid volume rate delivered by the pump"
- Create a new term MDC_FLOW_FLUID_PUMP_SETTING with the definition "current fluid volume rate setting" Note: This represents the current set rate with the pump either in standby or infusing
Decisions/Issues:- Need to look at whether it is an observation or a setting
- MDC_FLOW_FLUID_PUMP and MDC_FLOW_FLUID_PUMP_CURRENT end up being the same and will be used to report the current rate. _CURRENT has been used in the TF since 2019 and is used in some pump implementations.
- MDC_FLOW_FLUID_PUMP_SETTING will be used for PCD-03 to program and in PCD-01/10 to report what it is programmed to do. Added to containment model, based on MDC_FLOW_FLUID_PUMP with part = 258 instead of 2
Action(s):
3 PCD01 Containment
- KurtStatus/Discussion:
Decisions/Issues:- _MDC_ATTR_AL_COND_PUMP_GENERAL should be _MDC_ATTR_AL_COND for MDC_ATTR_AL_COND
- Missing enum values for _MDC_PUMP_DOSE_DELAY_INTERVAL_TYPE
Action(s):
4 Next Meeting
- ChairStatus/Discussion: Decisions/Issues:
- Topics will include ...
Action(s):
Next Meeting
NOTE: Next Meeting will be held during the regularly scheduled time: 12:00 (Eastern) on Jan 24 (60 minutes)