Line 38: | Line 38: | ||
:'''Chair/Host:''' ''Kurt Elliason (Smiths Medical)'' | :'''Chair/Host:''' ''Kurt Elliason (Smiths Medical)'' | ||
− | :'''Participants:''' '' | + | :'''Participants:''' '' Adam Bunton(Cerner), Al Engelbert (BBraun), Andrea Robertson (Smiths), Brian Sullivan(BD), Brian Witkowski (BBraun), Jeff Rinda (Interoperability Solutions), Paul Schluter (), Rachel Golan (Root3 Labs), Steve Dain(), Tom Kowalczyk (BBraun); |
+ | |||
Line 75: | Line 76: | ||
'''Decisions/Issues:''' | '''Decisions/Issues:''' | ||
− | + | :Need to look at whether it is an observation or a setting | |
+ | :MDC_FLOW_FLUID_PUMP and MDC_FLOW_FLUID_PUMP_CURRENT end up being the same and will be used to report the current rate. _CURRENT has been used in the TF since 2019 and is used in some pump implementations. | ||
+ | :MDC_FLOW_FLUID_PUMP_SETTING will be used for PCD-03 to program and in PCD-01/10 to report what it is programmed to do. Added to containment model, based on MDC_FLOW_FLUID_PUMP with part = 258 instead of 2 | ||
'''Action(s):''' | '''Action(s):''' | ||
Line 82: | Line 85: | ||
|- | |- | ||
| align="center" | 3 | | align="center" | 3 | ||
− | | ''' | + | | '''PCD01 Containment''' <br>- Kurt |
| '''Status/Discussion:''' | | '''Status/Discussion:''' | ||
'''Decisions/Issues:''' | '''Decisions/Issues:''' | ||
− | + | :_MDC_ATTR_AL_COND_PUMP_GENERAL should be _MDC_ATTR_AL_COND for MDC_ATTR_AL_COND | |
+ | : Missing enum values for _MDC_PUMP_DOSE_DELAY_INTERVAL_TYPE | ||
'''Action(s):''' | '''Action(s):''' |
Latest revision as of 13:46, 10 January 2022
Meeting Purpose
IHE PCD Infusion Pump Working Group discussions.
WebEx Information
Topic: IHE PCD Infusion Pump WG
Date: Monday, Jan 10, 2022
Time: 12:00, Eastern Time (GMT -05:00, New York)
Duration: 60 Minutes
Note: Specific web & phone information will be provided via e-mail to group members.
Contact Kurt Elliason for more information.
Proposed Agenda
- 1. Review today's Agenda
- 2. Review 2021-12-13 Pump Zoom Session
- 3. Discussion of MDC_FLOW_FLUID_PUMP, MDC_FLOW_FLUID_PUMP_CURRENT, MDC_FLOW_FLUID_PUMP_SETTING
- 4. Review containment draft for pcd-01 - enumerations
- 5. Review PCD Device Infusion Pump Action Items
- 6. New Business / Next Meetings
Attachments / Materials
- <add attachments links here>
- PCD03 containment for discussion
Discussion Notes
- Chair/Host: Kurt Elliason (Smiths Medical)
- Participants: Adam Bunton(Cerner), Al Engelbert (BBraun), Andrea Robertson (Smiths), Brian Sullivan(BD), Brian Witkowski (BBraun), Jeff Rinda (Interoperability Solutions), Paul Schluter (), Rachel Golan (Root3 Labs), Steve Dain(), Tom Kowalczyk (BBraun);
- NOTE: Action Item Review updates have been posted directly to the Infusion Pump Action Items wiki page
Item Topic Discussion 1 Agenda & Discussion Notes Review
- ChairStatus/Discussion:
Decisions/Issues:
Action(s):2 MDC_FLOW_FLUID_PUMP
- Chair + Steve DainStatus/Discussion: - Existing term MDC_FLOW_FLUID_PUMP 26712 (secondary refid as MDC_RATE_INFUS) which is currently defined as "Volume rate delivered by pump" in the 10101:2019 and is defined differently as "Programmed volumetric delivery rate" in latest PCD10 containment spreadsheet
- New proposed term MDC_FLOW_FLUID_PUMP_CURRENT code 26942 defined as "current pump fluid flow"
- Historically we have used 26712 as the flow of the pump.
- Pump WG desired to communicate what the pump rate is set at, even when the pump is not infusing.
- We may have created some difficulties with the proposed term and the "new" definition of an old term.
- Suggestion is to delete the proposed term MDC_FLOW_FLUID_PUMP_CURRENT, clarifying the definition of MDC_FLOW_FLUID_PUMP 26712 as "current fluid volume rate delivered by the pump"
- Create a new term MDC_FLOW_FLUID_PUMP_SETTING with the definition "current fluid volume rate setting" Note: This represents the current set rate with the pump either in standby or infusing
Decisions/Issues:- Need to look at whether it is an observation or a setting
- MDC_FLOW_FLUID_PUMP and MDC_FLOW_FLUID_PUMP_CURRENT end up being the same and will be used to report the current rate. _CURRENT has been used in the TF since 2019 and is used in some pump implementations.
- MDC_FLOW_FLUID_PUMP_SETTING will be used for PCD-03 to program and in PCD-01/10 to report what it is programmed to do. Added to containment model, based on MDC_FLOW_FLUID_PUMP with part = 258 instead of 2
Action(s):
3 PCD01 Containment
- KurtStatus/Discussion:
Decisions/Issues:- _MDC_ATTR_AL_COND_PUMP_GENERAL should be _MDC_ATTR_AL_COND for MDC_ATTR_AL_COND
- Missing enum values for _MDC_PUMP_DOSE_DELAY_INTERVAL_TYPE