Proposed Agenda
- 1. Review today's Agenda
- 2. Review 2021-05-24 Pump WebEx Session
- 3. MDCX terms
- 4a. Units of measure (mg of X)
- 4b. Priming Info Channel
- 5. Review PCD Device Infusion Pump Action Items
- 6. New Business / Next Meetings
Attachments / Materials
- <add attachments links here>
- proposed containment model
- pump enums
- PCD03 terms
Discussion Notes
- Chair/Host: Kurt Elliason (Smiths Medical)
- Participants: Al Engelbert (BBraun), Andrea Robertson (Smiths), Brian Sullivan(BD), Jeff Rinda (Interoperability Solutions), Marufur Rahmen (ICU), Michael Workman (BD), Paul Schluter (), Tom Kowalczyk (BBraun);
- NOTE: Action Item Review updates have been posted directly to the Infusion Pump Action Items wiki page
Item Topic Discussion 1 Agenda & Discussion Notes Review
- ChairStatus/Discussion:
Decisions/Issues:
Action(s):2 Planned topics
- tbdStatus/Discussion: - MDC_VOL_PCA_DOSE_LIMIT is missing from RTMMS. CF_CODE10 was defined along with several others, the ones before and after this one are in RTMMS. Need to find original description and get this one added.
- Jeff has descriptions for other MDCX terms.
- MDCX_INFUS_ORDER_STEP_NUM
- MDCX_INFUS_TOTAL_NUM_STEPS
- MDCX_INFUS_ORDER_STEP_TYPE - also needs new enum defined _MDC_INFUS_ORDER_STEP_TYPE
- MDCX_INFUSION_ORDER_TYPE - also needs new enum defined _MDC_INFUS_ORDER_TYPE
- Jeff has descriptions for other MDCX terms.
- Units of measure (mg of X) - did not discuss today
- Priming Info channel
- Discussed draft proposal. :proposed containment model how does this info channel interact with data in other channels.
- If Priming Info channel includes MDC_FLOW_FLUID_PUMP How is this different from MDC_FLOW_FLUID_PUMP_CURRENT in the deliv channel?
- Might need three new terms
- MDC_VOL_FLUID_PRIMING
- MDC_VOL_FLUID_PRIMING_REMAIN
- MDC_VOL_FLUID_PRIMING_TBI
Decisions/Issues:
Action(s):
3 TF Vol 2 question
- tbdStatus/Discussion: - Need to correct table in TF Vol 2 section 3.3.4.4.10.2
- Heading in Col 7 should read
- OBX segment with OBX-3 = MDCX_INFUSION_ ORDER_TYPE w/OBX-5 = clinician-dose
- Heading in Col 8 should read
- OBX segment with OBX-3 = MDC_FLOW_ FLUID_PUMP
- Heading in Col 7 should read
- Second bullet in "OBX segment" section near line 1070 should be changed to
- OBX-3 = MDCX_INFUSION_ORDER_TYPE. This term has enumerations in OBX-5 of “clinician-dose”, “loading-dose”, or “continuous”.
Decisions/Issues:
Action(s):
4 Next Meeting
- ChairStatus/Discussion: Decisions/Issues:
- Topics will include ...
Action(s):
Next Meeting
NOTE: Next Meeting will be held during the regularly scheduled time: 11:00 (Eastern) on June 28 (60 minutes)
(Reviewed & approved by PCD Infusion Pump WG on <TBD>