PCD Pump 2020-12-14 WebEx
Jump to navigation
Jump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.
Meeting Purpose
IHE PCD Infusion Pump Working Group discussions.
WebEx Information
Topic: IHE PCD Infusion Pump WG
Date: Monday, December 14, 2020
Time: 12:00, Eastern Time (GMT -05:00, New York)
Duration: 60 Minutes
Note: Specific web & phone information will be provided via e-mail to group members.
Contact Kurt Elliason for more information.
Proposed Agenda
- 1. Review today's Agenda
- 2. Review 2020-11-23 Pump WebEx Session
- 3. Containment tree updates
- 4.
- 5. Review PCD Device Infusion Pump Action Items
- 6. New Business / Next Meetings
Attachments / Materials
- <add attachments links here>
Discussion Notes
- Chair/Host: Kurt Elliason (Smiths Medical)
- Participants: Al Engelbert (BBraun), Andrea Robertson (Smiths), Brian Witkowski (BBraun), Jeff Rinda (Interoperability Solutions), Marufur Rahmen (ICU), Paul Schluter(CMI), Tom Kowalczyk (BBraun)
- NOTE: Action Item Review updates have been posted directly to the Infusion Pump Action Items wiki page
Item Topic Discussion 1 Agenda & Discussion Notes Review
- ChairStatus/Discussion:
Decisions/Issues:
Action(s):2 tbd
- tbdStatus/Discussion: - Discussed Rack and position
- Attributes are defined in Vol 3 for VMD
- MDC_ATTR_ID_POSN
- Bbraun has Pillar/Tower-Rack-Slot Pillars/Towers of up to 6 racks, racks contain up to 4 slots
- Fresenius use this model to convey which rack and slot a pump is physically in.
- OBX|1||69985^MDC_DEV_PUMP_INFUS_MDS^MDC|1.0.0.0|||||||X|||||||178864/21450198&Left&4^Fresenius^0013380000000000^EUI-64
OBX|2||70054^MDC_DEV_PUMP_INFUS_SYRINGE_VMD^MDC|1.1.0.0|||||||X|||||||Z018792/23225211&1^Fresenius^0013380000000000^EUI-64
- In OBX1, we identify the rack’s type and serial number and rack’s position(Left) and rack’s capacity(4)
- In OBX2, we identify the pump’s type and serial number and the pump’s position on the rack(1)
- Do we need new terms or just use a string? Need to make it scalable.
- Need to do some research into how to identify specific position of a pump. Probably need a CP for Vol 3, maybe 2.
- Discussion of Connectathon21, planning to join Connectathon call tomorrow
Decisions/Issues:
Action(s):- Add to January 11th agenda
- Discussion of FHIR based approach to pump programming - Brian W.
- CP for position in Pillar/Tower-Rack-Slot
- CP to document which terms are under the MDS if it is a Rack and which terms are under the pump's VMD. Provide a picture.
3 tbd
- tbdStatus/Discussion:
Decisions/Issues:
Action(s):
4 Next Meeting
- ChairStatus/Discussion: Decisions/Issues:
- Topics will include ...
Action(s):
- Discussed Rack and position
Next Meeting
NOTE: Next Meeting will be held during the regularly scheduled time: 11:00 (Eastern) on January 11 (60 minutes)
(Reviewed & approved by PCD Infusion Pump WG on <TBD>