Line 64: | Line 64: | ||
| '''Status/Discussion:''' | | '''Status/Discussion:''' | ||
+ | :Discussed Rack and position | ||
+ | ::Attributes are defined in Vol 3 for VMD | ||
+ | ::MDC_ATTR_ID_POSN | ||
+ | |||
+ | ::Bbraun has Pillar/Tower-Rack-Slot Pillars/Towers of up to 6 racks, racks contain up to 4 slots | ||
+ | |||
+ | ::Fresenius use this model to convey which rack and slot a pump is physically in. | ||
+ | ::OBX|1||69985^MDC_DEV_PUMP_INFUS_MDS^MDC|1.0.0.0|||||||X|||||||178864/21450198&Left&4^Fresenius^0013380000000000^EUI-64 | ||
+ | OBX|2||70054^MDC_DEV_PUMP_INFUS_SYRINGE_VMD^MDC|1.1.0.0|||||||X|||||||Z018792/23225211&1^Fresenius^0013380000000000^EUI-64 | ||
+ | |||
+ | ::In OBX1, we identify the rack’s type and serial number and rack’s position(Left) and rack’s capacity(4) | ||
+ | |||
+ | ::In OBX2, we identify the pump’s type and serial number and the pump’s position on the rack(1) | ||
+ | |||
+ | ::Do we need new terms or just use a string? Need to make it scalable. | ||
+ | |||
+ | ::Need to do some research into how to identify specific position of a pump. Probably need a CP for Vol 3, maybe 2. | ||
+ | |||
+ | |||
+ | :Discussion of Connectathon21, planning to join Connectathon call tomorrow | ||
'''Decisions/Issues:''' | '''Decisions/Issues:''' | ||
Line 69: | Line 89: | ||
'''Action(s):''' | '''Action(s):''' | ||
+ | :Add to January 11th agenda | ||
+ | ::Discussion of FHIR based approach to pump programming - Brian W. | ||
+ | ::CP for position in Pillar/Tower-Rack-Slot | ||
+ | ::CP to document which terms are under the MDS if it is a Rack and which terms are under the pump's VMD. Provide a picture. | ||
Latest revision as of 18:25, 16 December 2020
Meeting Purpose
IHE PCD Infusion Pump Working Group discussions.
WebEx Information
Topic: IHE PCD Infusion Pump WG
Date: Monday, December 14, 2020
Time: 12:00, Eastern Time (GMT -05:00, New York)
Duration: 60 Minutes
Note: Specific web & phone information will be provided via e-mail to group members.
Contact Kurt Elliason for more information.
Proposed Agenda
- 1. Review today's Agenda
- 2. Review 2020-11-23 Pump WebEx Session
- 3. Containment tree updates
- 4.
- 5. Review PCD Device Infusion Pump Action Items
- 6. New Business / Next Meetings
Attachments / Materials
- <add attachments links here>
Discussion Notes
- Chair/Host: Kurt Elliason (Smiths Medical)
- Participants: Al Engelbert (BBraun), Andrea Robertson (Smiths), Brian Witkowski (BBraun), Jeff Rinda (Interoperability Solutions), Marufur Rahmen (ICU), Paul Schluter(CMI), Tom Kowalczyk (BBraun)
- NOTE: Action Item Review updates have been posted directly to the Infusion Pump Action Items wiki page
Item Topic Discussion 1 Agenda & Discussion Notes Review
- ChairStatus/Discussion:
Decisions/Issues:
Action(s):2 tbd
- tbdStatus/Discussion: - Discussed Rack and position
- Attributes are defined in Vol 3 for VMD
- MDC_ATTR_ID_POSN
- Bbraun has Pillar/Tower-Rack-Slot Pillars/Towers of up to 6 racks, racks contain up to 4 slots
- Fresenius use this model to convey which rack and slot a pump is physically in.
- OBX|1||69985^MDC_DEV_PUMP_INFUS_MDS^MDC|1.0.0.0|||||||X|||||||178864/21450198&Left&4^Fresenius^0013380000000000^EUI-64
OBX|2||70054^MDC_DEV_PUMP_INFUS_SYRINGE_VMD^MDC|1.1.0.0|||||||X|||||||Z018792/23225211&1^Fresenius^0013380000000000^EUI-64
- In OBX1, we identify the rack’s type and serial number and rack’s position(Left) and rack’s capacity(4)
- In OBX2, we identify the pump’s type and serial number and the pump’s position on the rack(1)
- Do we need new terms or just use a string? Need to make it scalable.
- Need to do some research into how to identify specific position of a pump. Probably need a CP for Vol 3, maybe 2.
- Discussion of Connectathon21, planning to join Connectathon call tomorrow
Decisions/Issues:
Action(s):- Add to January 11th agenda
- Discussion of FHIR based approach to pump programming - Brian W.
- CP for position in Pillar/Tower-Rack-Slot
- CP to document which terms are under the MDS if it is a Rack and which terms are under the pump's VMD. Provide a picture.
3 tbd
- tbdStatus/Discussion:
Decisions/Issues:
Action(s):
4 Next Meeting
- ChairStatus/Discussion: Decisions/Issues:
- Topics will include ...
Action(s):
- Discussed Rack and position
Next Meeting
NOTE: Next Meeting will be held during the regularly scheduled time: 11:00 (Eastern) on January 11 (60 minutes)
(Reviewed & approved by PCD Infusion Pump WG on <TBD>