Proposed Agenda

1. Review today's Agenda
2. Review 2020-05-11 Pump WebEx Session
3. Review PCA Door event terms
4. Review IPEC tests
5. Review PCD Device Infusion Pump Action Items
6. New Business / Next Meetings

Attachments / Materials

  • <add attachments links here>
Working copy of IPEC tests https://drive.google.com/file/d/1_PVitRThfnAxms0P7OhnfRrbEozjBu3n/view?usp=sharing

Discussion Notes

Chair/Host: Kurt Elliason (Smiths Medical)
Participants: Al Engelbert (BBraun), Bill Haralson (Smiths), Christophe Fournier (Fresenius), Jeff Rinda (ICU Medical), Marufur Rahmen (ICU), Paul Schluter (CMI), Sarah Johnston (BBraun), Tom Kowalczyk (BBraun)
NOTE: Action Item Review updates have been posted directly to the Infusion Pump Action Items wiki page


Item Topic Discussion
1 Agenda & Discussion Notes Review
- Chair
Status/Discussion:


Decisions/Issues:


Action(s):

2 PCA Door Events
- Kurt
Status/Discussion:


Decisions/Issues: MDC_EVT_DROP_SENSOR_DISCONNECTION_REQD - drops of liquid. Sensor needs to be disconnected due to incompatibility with the selected drug. Needs more consideration.

MDC_EVT_PCA_INSUF_DOSE should be MDC_EVT_TCI_INSUF_DOSE Decided to use MDC_EVT_TCI _VOL_LOW - In TCI mode, the remaining volume/dose in the syringe is insufficient to reach the target - Not enough medication in the syringe to achieve the desired target.

Change MDC_EVT_PCA_MAX_BOLUSES to MDC_EVT_PCA_MAX_BOLUS_COUNT_EXCEEDED   Having the PCA_NEAR_MAX_LIMIT companion to PCA_MAX_LIMIT aligns other term sets where we have both and “ending” and near “ending” events.

Action(s):


3 IPEC Test updates
- Kurt
Status/Discussion:


Decisions/Issues: Final version ready for PCD review during Connectathon test review meeting. https://drive.google.com/file/d/13QapYk0Air5WYsWMr65k-Hvun9770Mlg/view?usp=sharing