Difference between revisions of "PCD Pump 2020-06-08 WebEx"
(2 intermediate revisions by the same user not shown) | |||
Line 32: | Line 32: | ||
:* '''<add attachments links here>''' | :* '''<add attachments links here>''' | ||
− | : IPEC tests https://drive.google.com/ | + | : Working copy of IPEC tests https://drive.google.com/file/d/1_PVitRThfnAxms0P7OhnfRrbEozjBu3n/view?usp=sharing |
== Discussion Notes == | == Discussion Notes == | ||
Line 38: | Line 38: | ||
:'''Chair/Host:''' ''Kurt Elliason (Smiths Medical)'' | :'''Chair/Host:''' ''Kurt Elliason (Smiths Medical)'' | ||
− | :'''Participants:''' '' | + | :'''Participants:''' '' Al Engelbert (BBraun), Bill Haralson (Smiths), Christophe Fournier (Fresenius), Jeff Rinda (ICU Medical), Marufur Rahmen (ICU), Paul Schluter (CMI), Sarah Johnston (BBraun), Tom Kowalczyk (BBraun)'' |
::'''NOTE:''' '''''Action Item Review updates have been posted directly to the [[PCD_Device_Infusion_Pump_Action_Items | Infusion Pump Action Items wiki page]]''''' | ::'''NOTE:''' '''''Action Item Review updates have been posted directly to the [[PCD_Device_Infusion_Pump_Action_Items | Infusion Pump Action Items wiki page]]''''' | ||
Line 61: | Line 61: | ||
|- | |- | ||
| align="center" | 2 | | align="center" | 2 | ||
− | | ''' | + | | '''PCA Door Events''' <br>- Kurt |
| '''Status/Discussion:''' | | '''Status/Discussion:''' | ||
'''Decisions/Issues:''' | '''Decisions/Issues:''' | ||
+ | MDC_EVT_DROP_SENSOR_DISCONNECTION_REQD - drops of liquid. Sensor needs to be disconnected due to incompatibility with the selected drug. Needs more consideration. | ||
+ | MDC_EVT_PCA_INSUF_DOSE should be MDC_EVT_TCI_INSUF_DOSE | ||
+ | Decided to use | ||
+ | MDC_EVT_TCI _VOL_LOW - In TCI mode, the remaining volume/dose in the syringe is insufficient to reach the target | ||
+ | - Not enough medication in the syringe to achieve the desired target. | ||
+ | |||
+ | Change MDC_EVT_PCA_MAX_BOLUSES to MDC_EVT_PCA_MAX_BOLUS_COUNT_EXCEEDED | ||
+ | |||
+ | Having the PCA_NEAR_MAX_LIMIT companion to PCA_MAX_LIMIT aligns other term sets where we have both and “ending” and near “ending” events. | ||
'''Action(s):''' | '''Action(s):''' | ||
Line 73: | Line 82: | ||
|- | |- | ||
| align="center" | 3 | | align="center" | 3 | ||
− | | ''' | + | | '''IPEC Test updates''' <br>- Kurt |
| '''Status/Discussion:''' | | '''Status/Discussion:''' | ||
'''Decisions/Issues:''' | '''Decisions/Issues:''' | ||
− | + | Final version ready for PCD review during Connectathon test review meeting. | |
+ | https://drive.google.com/file/d/13QapYk0Air5WYsWMr65k-Hvun9770Mlg/view?usp=sharing | ||
'''Action(s):''' | '''Action(s):''' |
Latest revision as of 12:43, 8 June 2020
Meeting Purpose
IHE PCD Infusion Pump Working Group discussions.
WebEx Information
Topic: IHE PCD Infusion Pump WG
Date: Monday, June 8, 2020
Time: 11:00, Eastern Time (GMT -05:00, New York)
Duration: 60 Minutes
Note: Specific web & phone information will be provided via e-mail to group members.
Contact Kurt Elliason for more information.
Proposed Agenda
- 1. Review today's Agenda
- 2. Review 2020-05-11 Pump WebEx Session
- 3. Review PCA Door event terms
- 4. Review IPEC tests
- 5. Review PCD Device Infusion Pump Action Items
- 6. New Business / Next Meetings
Attachments / Materials
- <add attachments links here>
- Working copy of IPEC tests https://drive.google.com/file/d/1_PVitRThfnAxms0P7OhnfRrbEozjBu3n/view?usp=sharing
Discussion Notes
- Chair/Host: Kurt Elliason (Smiths Medical)
- Participants: Al Engelbert (BBraun), Bill Haralson (Smiths), Christophe Fournier (Fresenius), Jeff Rinda (ICU Medical), Marufur Rahmen (ICU), Paul Schluter (CMI), Sarah Johnston (BBraun), Tom Kowalczyk (BBraun)
- NOTE: Action Item Review updates have been posted directly to the Infusion Pump Action Items wiki page
Item Topic Discussion 1 Agenda & Discussion Notes Review
- ChairStatus/Discussion:
Decisions/Issues:
Action(s):2 PCA Door Events
- KurtStatus/Discussion:
Decisions/Issues: MDC_EVT_DROP_SENSOR_DISCONNECTION_REQD - drops of liquid. Sensor needs to be disconnected due to incompatibility with the selected drug. Needs more consideration.MDC_EVT_PCA_INSUF_DOSE should be MDC_EVT_TCI_INSUF_DOSE Decided to use MDC_EVT_TCI _VOL_LOW - In TCI mode, the remaining volume/dose in the syringe is insufficient to reach the target - Not enough medication in the syringe to achieve the desired target.
Change MDC_EVT_PCA_MAX_BOLUSES to MDC_EVT_PCA_MAX_BOLUS_COUNT_EXCEEDED Having the PCA_NEAR_MAX_LIMIT companion to PCA_MAX_LIMIT aligns other term sets where we have both and “ending” and near “ending” events.
Action(s):
3 IPEC Test updates
- KurtStatus/Discussion:
Decisions/Issues: Final version ready for PCD review during Connectathon test review meeting. https://drive.google.com/file/d/13QapYk0Air5WYsWMr65k-Hvun9770Mlg/view?usp=sharingAction(s):
4 Next Meeting
- ChairStatus/Discussion: Decisions/Issues:
- Topics will include ...
Action(s):
Next Meeting
NOTE: Next Meeting will be held during the regularly scheduled time: 11:00 (Eastern) on June 22 (60 minutes)
(Reviewed & approved by PCD Infusion Pump WG on <TBD>