The '''''MR Diffusion Imaging (DIFF) Integration Profile''''':
 
The '''''MR Diffusion Imaging (DIFF) Integration Profile''''':
* Facilitates the exchange of enhanced MR diffusion images containing standard attributes for Diffusion b-value, Diffusion gradient orientation and Diffusion image type. These objects are exchanged in an interoperable way allowing correct and convenient display.  
+
* Facilitates the interoperable exchange and correct display of enhanced MR diffusion images containing standard attributes for Diffusion b-value, Diffusion gradient orientation and Diffusion image type.  
 +
* Creates stacks of images (slices) with different meaning and derived contrast by using techniques of MR Diffusion Weighted Imaging (DWI)or Diffusion Tensor Imaging (DTI)
 +
* Defines how a set of images resulting from a completed MR Diffusion exam are stored and archived into frames of one Enhanced MR DICOM (Multi-frame) object so that they can later be retrieved and displayed in an unambiguous way on image viewers supporting Enhanced MR DICOM objects.
 +
* Enables widespread and adequate review and interpretation of b=0, ISOTROPIC and ADC diffusion frames by radiologists in an environment where all Actors support the Enhanced MR DICOM objects
 +
* A complete set of original diffusion frames is acquired and derived frames like ADC images or ISOTROPIC images are created by the Evidence Creator actor. The derived images can be created either in the same activity of the original images creation or in a separate activity
 +
 
 +
 
  
* Defines how the perfusion images are stored and archived into frames of one Enhanced DICOM (Multi-frame) object so that they can later be retrieved and displayed in an unambiguous way on image viewers supporting Enhanced DICOM objects.
 
 
* The images need to be stored with indices along their position in the stack and the progress of time.
 
* The images need to be stored with indices along their position in the stack and the progress of time.
 
* A stack of images is being created at different timeslots, in order to display the contrast agent over time
 
* A stack of images is being created at different timeslots, in order to display the contrast agent over time

Revision as of 14:59, 25 February 2010

MR Diffusion Imaging (DIFF) supports the exchange and display of Enhanced MR diffusion images.

Summary

The MR Diffusion Imaging Integration Profile leverages the Enhanced MR diffusion images containing standard attributes for Diffusion b-value, Diffusion gradient orientation and Diffusion image type. These objects are exchanged in an interoperable way allowing correct and convenient display.

Perfusion Fig1.jpg

Benefits

Support the Enhanced MR procedures for standard diffusion

  • Stacks of Images (Slices) are generated with different meaning and derived contrast for:
- Diffusion Weighted Images (DWI)
- Diffusion Tensor Imaging (DTI)

Improves the Consistency and Interoperability of the Enhanced MR Disffusion, clinical applications

  • Provides standard attributes for Diffusion b-value, Diffusion gradient orientation and Diffusion image type.
  • Reduce implementations diversity of choices for improving interoperability

Improve display

  • Creates stacks of images at different timeslots, in order to demonstrate the uptake of a contrast agent over time.
  • Defines how the perfusion images are stored and archived into frames of one Enhanced DICOM (Multi-frame) object so that they can later be retrieved and displayed in an unambiguous way on image viewers supporting Enhanced DICOM objects

Improves the acceptance of the Enhanced MR SOP Classes by industry and users

Reduces Deployment Cost/Time

Details

The MR Diffusion Imaging (DIFF) Integration Profile:

  • Facilitates the interoperable exchange and correct display of enhanced MR diffusion images containing standard attributes for Diffusion b-value, Diffusion gradient orientation and Diffusion image type.
  • Creates stacks of images (slices) with different meaning and derived contrast by using techniques of MR Diffusion Weighted Imaging (DWI)or Diffusion Tensor Imaging (DTI)
  • Defines how a set of images resulting from a completed MR Diffusion exam are stored and archived into frames of one Enhanced MR DICOM (Multi-frame) object so that they can later be retrieved and displayed in an unambiguous way on image viewers supporting Enhanced MR DICOM objects.
  • Enables widespread and adequate review and interpretation of b=0, ISOTROPIC and ADC diffusion frames by radiologists in an environment where all Actors support the Enhanced MR DICOM objects
  • A complete set of original diffusion frames is acquired and derived frames like ADC images or ISOTROPIC images are created by the Evidence Creator actor. The derived images can be created either in the same activity of the original images creation or in a separate activity


  • The images need to be stored with indices along their position in the stack and the progress of time.
  • A stack of images is being created at different timeslots, in order to display the contrast agent over time
  • Defines the correct and convenient display of the Perfusion with Contrast
  • Requirements for the Perfusion Creator modality are:
 1. Create an Enhanced CT or MR  SOP Instance
 2. Modality provides Dimension indices: Stack ID, In Stack Position Number, Temporal Position Index
 3. Data requirements: Image and Frame Type should be: Value 1: ORIGINAL, Value 2: PRIMARY, Value 3: PERFUSION
  • Requirements for Image Display Perfusion with contrast:
 1. Provide at least one viewport for display of frames with the same  In-Stack Position Number (Slice-index). More viewports may be used for displaying images for other In-Stack Position Numbers= simultaneous scrolling
 2. Scroll by Temporal Position Index then Stack ID / In-Stack Position Number
    and by Stack ID / In-Stack Position Number then Temporal Position Index   
 3. Provide the time since the start, deducted from Frame Reference DateTime
 4. Display the values of the dimension attributes and contrast attributes
 5. Display Frame annotations for Frame Number, Slice offset, relative time.
Perfusion Fig2a.jpg

Applications using older single-frame DICOM image objects depended on knowledge of private attributes for this information.

Systems Affected

Systems involved in this profile are:

  • Image Display that Query / Retrieve the enhanced CT/MR images and displays the viewports of frames with the identical In-Stack Position Number (Slice Ingdex)
  • Radiology image manager/archive, Storing the images with indices along their position in the stack and the progress of time.
  • Acquisition modalities, creates the Enhanced CR or MR SOP instances with the appropirate dimension indices


Actors & Transactions: