==Summary==
 
==Summary==
  
The MR Diffusion Imaging Integration Profile leverages the Enhanced MR diffusion images containing standard attributes for Diffusion b-value, Diffusion gradient orientation and Diffusion image type. These objects are exchanged in an interoperable way allowing correct and convenient display.  
+
The Profile uses the DICOM Enhanced MR object to encode diffusion images.  Images are organized into stacks and labelled using standard attributes for Diffusion b-value, Diffusion gradient orientation and Diffusion image type.
 +
 
 +
By requiring the use of standard attributes (instead of variable private tags) Image Displays can reliably display diffusion imaging correctly and conveniently.  
 
   
 
   
<center>[[Image:Perfusion Fig1.jpg]]</center>
+
[[Image:Diff Dis.JPG|center|]]
  
 
==Benefits==
 
==Benefits==
Line 15: Line 17:
 
:- Diffusion Weighted Images (DWI)  
 
:- Diffusion Weighted Images (DWI)  
 
:- Diffusion Tensor Imaging (DTI)
 
:- Diffusion Tensor Imaging (DTI)
'''Improves the Consistency and Interoperability of the Enhanced MR Disffusion, clinical applications'''
+
'''Improves the Consistency and Interoperability of the Enhanced MR Diffusion, clinical applications'''
 
* Provides standard attributes for Diffusion b-value, Diffusion gradient orientation and Diffusion image type.  
 
* Provides standard attributes for Diffusion b-value, Diffusion gradient orientation and Diffusion image type.  
 
* Reduce implementations diversity of choices for improving interoperability
 
* Reduce implementations diversity of choices for improving interoperability
Line 33: Line 35:
 
* Defines how a set of images resulting from a completed MR Diffusion exam are stored and archived into frames of one Enhanced MR DICOM (Multi-frame) object so that they can later be retrieved and displayed in an unambiguous way on image viewers supporting Enhanced MR DICOM objects.
 
* Defines how a set of images resulting from a completed MR Diffusion exam are stored and archived into frames of one Enhanced MR DICOM (Multi-frame) object so that they can later be retrieved and displayed in an unambiguous way on image viewers supporting Enhanced MR DICOM objects.
 
* Enables widespread and adequate review and interpretation of b=0, ISOTROPIC and ADC diffusion frames by radiologists in an environment where all Actors support the Enhanced MR DICOM objects
 
* Enables widespread and adequate review and interpretation of b=0, ISOTROPIC and ADC diffusion frames by radiologists in an environment where all Actors support the Enhanced MR DICOM objects
* A complete set of original diffusion frames is acquired and derived frames like ADC images or ISOTROPIC images are created by the Evidence Creator actor. The derived images can be created either in the same activity of the original images creation or in a separate activity
+
* Functional requirements for Images Creator
 +
:1. Create a series of original images with different Diffusion b-values. Including a stack of images for b=0 and Images with other diffusion b value shell not be displayed
 +
:2. Create a series for resulting ISOTROPIC images where b=MAX
 +
:3. Create a series with ADC images for each slice and at least for b=MAX
 +
:4. The ISOTROPIC and ADC images for b=MAX shell be presented in conjunction with the slices that b=0
 +
* Data requirements for Images Creator: Stack ID, InStack position, Diffusion b-value
  
 +
<center>[[Image:Perfusion Fig1.jpg]]</center>
  
 +
* Functional requirements for Display Enhanced MR with Diffusion
 +
:1. Provide (at least) one row of three viewports for the display of frames for the same InStackPositionNumber / slice number
 +
:2. The frames for one InStackPositionNumber need to be presented in one row. Additional rows can be used for other InStackPositionNumber-s
 +
:3. Capable to filter / select  frames based  on Diffusion Directionality  attribute 
 +
:4. Capable to scroll the images (vertically) by InStackPositionNumber-Index
  
* The images need to be stored with indices along their position in the stack and the progress of time.
+
* MR Diffusion Imaging Process Flow
* A stack of images is being created at different timeslots, in order to display the contrast agent over time
+
: The derived images can be created either in the same activity of the original images creation or in a separate activity
* Defines the correct and convenient display of the Perfusion with Contrast
+
:1. Creating original and derived images as one activity - Process Flow
* Requirements for the Perfusion Creator modality are:
+
[[Image:Diff Flow1.JPG]]
  1. Create an Enhanced CT or MR  SOP Instance
+
:2. Creating derived images in a separate activity - Process Flow
  2. Modality provides Dimension indices: Stack ID, In Stack Position Number, Temporal Position Index
 
  3. Data requirements: Image and Frame Type should be: Value 1: ORIGINAL, Value 2: PRIMARY, Value 3: PERFUSION
 
* Requirements for Image Display Perfusion with contrast:
 
  1. Provide at least one viewport for display of frames with the same  In-Stack Position Number (Slice-index). More viewports may be used for displaying images for other In-Stack Position Numbers= simultaneous scrolling
 
  2. Scroll by Temporal Position Index then Stack ID / In-Stack Position Number
 
    and by Stack ID / In-Stack Position Number then Temporal Position Index 
 
  3. Provide the time since the start, deducted from Frame Reference DateTime
 
  4. Display the values of the dimension attributes and contrast attributes
 
  5. Display Frame annotations for Frame Number, Slice offset, relative time.
 
  
<center>[[Image:Perfusion Fig2a.jpg]]</center>
+
[[Image:Diff Flow2.JPG]]
Applications using older single-frame DICOM image objects depended on knowledge of private attributes for this information.
 
  
 
==Systems Affected==
 
==Systems Affected==
Line 59: Line 62:
 
Systems involved in this profile are:
 
Systems involved in this profile are:
  
* Image Display that Query / Retrieve the enhanced CT/MR images and displays the viewports of frames with the identical In-Stack Position Number (Slice Ingdex)
+
* Acquisition modalities create the Enhanced MR images with the appropriate diffusion data
* Radiology image manager/archive, Storing the images with indices along their position in the stack and the progress of time.
+
* Processing workstations can retrieve the Enhanced MR images and create Derived Images
* Acquisition modalities, creates the Enhanced CR or MR SOP instances with the appropirate dimension indices
+
* PACS systems store the enhanced images.
 +
* Image Displays retrieve the original and derived enhanced MR images and displays them
  
 +
'''Actors & Transactions:'''
  
'''Actors & Transactions:'''
+
[[Image:Diff Diag1.JPG]]
  
[[Image:Perfusion Fig4.JPG]]
 
 
==Specification==
 
==Specification==
  
Line 72: Line 76:
  
 
'''Documents:'''  
 
'''Documents:'''  
:* [http://www.ihe.net/Technical_Framework/upload/IHE-RAD_TF_Suppl_PERF_2009-06-21.pdf CT/MR Perfusion with Contrast Supplement]
+
:* [http://www.ihe.net/Technical_Framework/upload/IHE-RAD_TF_Suppl_DIFF_2009-06-21.pdf MR Diffusion Imaging (DIFF) Supplement]
  
 
'''Underlying Standards:'''
 
'''Underlying Standards:'''
Line 80: Line 84:
  
 
'''Related Profiles'''
 
'''Related Profiles'''
 +
* [[Scheduled Workflow]] [SWF] can manage the production of MR Diffusion Images.
 +
 +
* [[Patient Information Reconciliation]] [PIR] is expected to reconcile MR Diffusion Images along with the rest of the patient data record.
 +
 +
* [[Consistent Presentation of Images]] [CPI] can facilitate the Consistent Display and Print of MR Diffusion Images
 +
 +
* [[Reporting Workflow]] [RWF] may use MR Diffusion Images as inputs to the reporting process.
 +
 +
* [[Cross-enterprise Document Sharing for Imaging]] [XDS-I] can be used to share MR Diffusion Images between sites over a network.
 +
 +
* [[Portable Data for Imaging]] [PDI] can store MR Diffusion Images on media such as CDs.
 +
 +
* [[Import Reconciliation Workflow]] [IRWF] can fix patient ids, etc. of MR Diffusion Images when importing.
 +
 +
* [[Radiation Exposure Monitoring]] [REM] can record radiation details associated with the creation of the MR Diffusion Images.
  
  
Line 94: Line 113:
  
 
[[Category:Profiles]]
 
[[Category:Profiles]]
 +
[[Category:RAD Profile]]
 +
[[Category:DICOM]]

Latest revision as of 15:11, 25 May 2018

MR Diffusion Imaging (DIFF) supports the exchange and display of Enhanced MR diffusion images.

Summary

The Profile uses the DICOM Enhanced MR object to encode diffusion images. Images are organized into stacks and labelled using standard attributes for Diffusion b-value, Diffusion gradient orientation and Diffusion image type.

By requiring the use of standard attributes (instead of variable private tags) Image Displays can reliably display diffusion imaging correctly and conveniently.

Diff Dis.JPG

Benefits

Support the Enhanced MR procedures for standard diffusion

  • Stacks of Images (Slices) are generated with different meaning and derived contrast for:
- Diffusion Weighted Images (DWI)
- Diffusion Tensor Imaging (DTI)

Improves the Consistency and Interoperability of the Enhanced MR Diffusion, clinical applications

  • Provides standard attributes for Diffusion b-value, Diffusion gradient orientation and Diffusion image type.
  • Reduce implementations diversity of choices for improving interoperability

Improve display

  • Creates stacks of images at different timeslots, in order to demonstrate the uptake of a contrast agent over time.
  • Defines how the perfusion images are stored and archived into frames of one Enhanced DICOM (Multi-frame) object so that they can later be retrieved and displayed in an unambiguous way on image viewers supporting Enhanced DICOM objects

Improves the acceptance of the Enhanced MR SOP Classes by industry and users

Reduces Deployment Cost/Time

Details

The MR Diffusion Imaging (DIFF) Integration Profile:

  • Facilitates the interoperable exchange and correct display of enhanced MR diffusion images containing standard attributes for Diffusion b-value, Diffusion gradient orientation and Diffusion image type.
  • Creates stacks of images (slices) with different meaning and derived contrast by using techniques of MR Diffusion Weighted Imaging (DWI)or Diffusion Tensor Imaging (DTI)
  • Defines how a set of images resulting from a completed MR Diffusion exam are stored and archived into frames of one Enhanced MR DICOM (Multi-frame) object so that they can later be retrieved and displayed in an unambiguous way on image viewers supporting Enhanced MR DICOM objects.
  • Enables widespread and adequate review and interpretation of b=0, ISOTROPIC and ADC diffusion frames by radiologists in an environment where all Actors support the Enhanced MR DICOM objects
  • Functional requirements for Images Creator
1. Create a series of original images with different Diffusion b-values. Including a stack of images for b=0 and Images with other diffusion b value shell not be displayed
2. Create a series for resulting ISOTROPIC images where b=MAX
3. Create a series with ADC images for each slice and at least for b=MAX
4. The ISOTROPIC and ADC images for b=MAX shell be presented in conjunction with the slices that b=0
  • Data requirements for Images Creator: Stack ID, InStack position, Diffusion b-value
Perfusion Fig1.jpg
  • Functional requirements for Display Enhanced MR with Diffusion
1. Provide (at least) one row of three viewports for the display of frames for the same InStackPositionNumber / slice number
2. The frames for one InStackPositionNumber need to be presented in one row. Additional rows can be used for other InStackPositionNumber-s
3. Capable to filter / select frames based on Diffusion Directionality attribute
4. Capable to scroll the images (vertically) by InStackPositionNumber-Index
  • MR Diffusion Imaging Process Flow
The derived images can be created either in the same activity of the original images creation or in a separate activity
1. Creating original and derived images as one activity - Process Flow

Diff Flow1.JPG

2. Creating derived images in a separate activity - Process Flow

Diff Flow2.JPG

Systems Affected

Systems involved in this profile are:

  • Acquisition modalities create the Enhanced MR images with the appropriate diffusion data
  • Processing workstations can retrieve the Enhanced MR images and create Derived Images
  • PACS systems store the enhanced images.
  • Image Displays retrieve the original and derived enhanced MR images and displays them

Actors & Transactions:

Diff Diag1.JPG

Specification

Profile Status: Trial Implementation

Documents:

Underlying Standards:

See Also

Related Profiles


Consumer Information


Implementer Information


Reference Articles


This page is based on the Profile Template