Difference between revisions of "MESA/Dose Register"

From IHE Wiki
Jump to navigation Jump to search
(Created page with "<big>'''MESA/Dose Information Consumer'''</big> __TOC__ =Radiation Exposure Monitoring Integration Profile Tests= This section describes tests that are specific to the IHE R...")
 
 
(5 intermediate revisions by the same user not shown)
Line 1: Line 1:
<big>'''MESA/Dose Information Consumer'''</big>
+
REM Dose Register Preparatory tests have been moved to https://gazelle.ihe.net/content/preparatory-tests
__TOC__
 
 
 
=Radiation Exposure Monitoring Integration Profile Tests=
 
 
 
This section describes tests that are specific to the IHE Radiation Exposure Monitoring Integration (REM) profile.
 
 
 
==Dose Register Test 4400: REM Vendor Interoperability==
 
 
 
Dose Registers in the REM profile are required to support storage of DICOM Dose SR objects.
 
 
 
This test gives you access to sample DICOM SRs created by REM Acquisition Modalities prior to the current Connectathon.  Modalities are asked to submit samples into gazelle 2 weeks prior to the Connectathon test deadline (or earlier).  The goal is to avoid surprises on the Connectathon floor.
 
 
 
'''Instructions'''
 
 
 
*You will find these samples in gazelle under '''Connectathon -> List of Samples'''.  Select your system name and look under '''Samples available for rendering''' to find sample REM Dose SRs. 
 
 
 
*Your task is to test with these samples prior to the Connectathon.
 
 
 
'''Evaluation'''
 
 
 
*Create a short txt file indicating you have accessed these samples. List those that you have successfully tested with your Dose Register. If you encounter problems, contact the Project Manager who will help resolve them with Modality vendors. Upload '''that''' txt file into gazelle as the results for this test.
 
 
 
==Dose Register Test 4401: De-identified Dose SR==
 
 
 
This test is similar to test 4400.  It gives you access to sample DICOM SRs which have been '''de-identified''' by Dose Information Reporter systems.  These systems are asked to submit samples into gazelle 2 weeks prior to the preConnectathon test deadline (or earlier).  The goal is to avoid surprises on the Connectathon floor.
 
 
 
'''Instructions'''
 
 
 
*You will find these samples in gazelle under '''Connectathon -> List of Samples'''.  Select your system name and look under '''Samples available for rendering''' to find sample de-identified REM Dose SRs. 
 
 
 
*Your task is to test with these samples prior to the Connectathon.
 
 
 
'''Evaluation'''
 
 
 
*Create a short txt file indicating you have accessed these samples. List those that you have successfully tested with your Dose Register. If you encounter problems, contact the Project Manager who will help resolve them with Modality vendors. Upload '''that''' txt file into gazelle as the results for this test.
 
 
 
== Dose Register Test 4404: Access REM Samples from Past Connectathons ==
 
 
 
We have been testing REM for several years, and we've collected sample Dose SRs from Modalities that have tested REM at various connectathons.  To expand the number of samples you have to test with, we point you to this collection of SRs.
 
 
 
'''INSTRUCTIONS'''
 
 
 
* Access samples from past Connectathons stored here:  ftp://ftp.ihe.net/Connectathon/samples/RAD-profiles/REM_samples/  Note that these samples are as they were submitted by the modality vendor. They may or may not pass validation.
 
 
 
* We encourage you to download them and test with them in your lab.  At the Connectathon, you will be required to test with actual test partners, but we hope these samples provide for expanded testing opportunities.
 
 
 
'''EVALUATION'''
 
 
 
There is no log to upload for this test.
 

Latest revision as of 16:09, 26 July 2023

REM Dose Register Preparatory tests have been moved to https://gazelle.ihe.net/content/preparatory-tests