'''Basic Image Review (BIR)''' defines a set of baseline features and user interface relevant to the simple review of DICOM images.
  
'''To use this page programmatically, create a new profile description page named "Your Profile Name ''Profile'' and include the following line inside it:'''
+
__TOC__
 +
 
 +
==Summary==
 +
 
 +
Compliant software must provide a predictable user interface and functionality sufficient to review images for the purpose of clinical decision-making by ordering physicians:
 +
:* display of grayscale and color images from any modality
 +
:* visual navigation of the available series of images through the use of thumbnails
 +
:* side-by-side comparison of at least two sets of images (with synchronized scroll, pan and zoom for cross-sectional modalities)
 +
:* annotation of laterality, orientation, and spatial localization
 +
:* annotation of demographics, management and basic technique information for safe identification and usage
 +
:* simple measurements of linear distance and angle
 +
:* cine capability for images that involve cardiac motion (e.g., cardiac US, XA, 500 CT or MR)
  
<pre>
+
BIR is not intended to be a reading workstation.
{{subst::Profile Overview Template|
 
One line description |
 
One Paragraph Description |
 
Basic Image Review |
 
Radiology Technical Committee }}
 
</pre>
 
  
''' Save the page, and the boiler-plate text will be filled in for you'''
+
==Benefits==
  
''This template is for the one or two page user-oriented overview of an IHE Profile that is in Final Text, Trial Implementation or perhaps Public Comment.  Delete text in italics and replace it with your material.  Don't forget to delete the double quotes too.''
+
'''Reduce Errors or Delays in Patient Care'''
 +
* Reduce the need to work with unfamiliar interfaces
 +
* Reduce CD data that is effectively inaccessible by making sure viewers are runnable and usable
 +
* Reduce interpretation problems by ensuring critical basic functions are present
  
''Your page name should simply be Full Profile Name with spaces, with capitals, without the acronym. e.g. Scheduled Workflow. so it provides a title to the page. You can redirect the acronym to the full named page if you like for bonus points.''
+
'''Improve Efficiency'''
 +
* Reduce time spent learning new viewer interfaces by requiring similar layout, icons and interactions
 +
* Reduce time wasted on viewers that do not run and images that do not load by requiring support for a basic reference platform and minimum installation footprint.
  
 +
'''Reduce Cost and Dose'''
 +
* Avoid repeating scans by making the original results properly accessible
  
''<IN ONE LINE, tell a user what the profile is about (including the acronym) so they can decide if they're on the right page.  Basically this should be the same sentence that appears on the [[Profiles]] catalog page>''
+
==Details==
  
''e.g. Scheduled Workflow (SWF) integrates ordering, scheduling, imaging acquisition, storage and viewing for Radiology exams.''
+
To ensure adequate, predictable functionality, the profile defines a specific, simplified set of functions.
  
}}}
+
To minimize errors and learning time, the user interface elements are described to the level needed for consistency, while allowing the implementer some flexibility with the fine details. 
__TOC__
 
  
==Summary==
+
The '''BIR user interface''' includes:
{{{2|''<Describe the profile in about a paragraph using user-oriented language.  Focus on what it accomplishes for a user (i.e. the Use Cases).  Don't get into how it works, leave that to the Details section.>''}}}
+
* a strip or panel containing a thumbnail for each series
 +
* one or more viewports which open to display selected series following predictable default image guidance
 +
* a toolbar or panel containing a standard set of icons for the '''specified functions'''
 +
** patient selection
 +
** image, series, study navigation within a patient
 +
** viewport and tile layout control
 +
** viewport selection and synchronization control
 +
** windowing and greyscale inversion control and presets
 +
** scrolling, rotation, panning and zooming control
 +
** annotation, localizer and crosshair display control
 +
** distance and angle measurement
 +
** cine control
 +
** report display control
 +
** printing
  
''<Insert a simple graphic that, at a glance, visually summarizes what the profile is about.  Do not use an actor/transaction diagram here.  Show your graphic to someone for 5 seconds (literally) and ask them what it's about.  If what they say hits the main points in your summary paragraph, you have succeeded.  E.g. a graphic of a hospital, a clinic, and a lab with patient records moving between them. .>''
+
The profile allows implementers to include an '''"Advanced" button''' which gives users the choice of moving past the initial familiar interface and gives vendors a place to demonstrate their creativity.
  
''<See [[Help:Contents#Tips_.26_Tricks| Help - Tips and Tricks]] for details on inserting an image/graphic.>''
+
The profile does not require (or permit) advanced image processing like pseudo-coloring, multi-planar reconstruction, multi-modality fusion, surface rendering, volume rendering, etc. in the basic interface.  Vendors may choose to provide such functionality in their Advanced interface.  
  
==Benefits==
+
An implementation may choose to support screen calibration such as that described in the [[Consistent Presentation of Images]] Profile however the profile does not mandate it since this often depends on hardware which may not be present on the system where the BIR software is being run.  As a result, the software may inform the user that the display environment may or may not be adequate for diagnostic purposes.  However to comply with the profile there shall be nothing about the software or the data which precludes '''diagnostic use''' so the software may NOT declare either to be unsuitable.
  
''<List the key benefits the profile provides (e.g. error reduction, increased throughput) and how they come about (e.g. SWF reduces patient errors due to mistyped demographics at the modality by transfering demographics electronically from the Order Filler). Consider using a bullet list for readability>''
+
Finally, the profile does not address application or modality specific requirements, such as may be needed for primary interpretation of mammography by radiologists, or primary interpretation of nuclear medicine image, which are addressed by other profiles.
  
==Details==
+
A '''Basic Review Option in [[Portable Data for Imaging]]''' specifies media creators be capable of including a BIR compliant viewer.
  
''<A few paragraphs, if appropriate, providing more details (mostly in user-speak, not tech-speak) on what the profile does and how it works.>''
+
To address installation issues, BIR viewers on PDI media are required to be able to run on a basic reference platform and install with a minimum footprint.
  
''<If the user might be familiar with the mechanisms used by the profile, you can mention them here.  E.g. Evidence Documents is based on DICOM Structured Report (SR) Templates.>''
+
To address performance issues, BIR viewers on PDI media are required to meet certain '''performance targets''' on the reference platform, e.g. take no more than 60 seconds to go from startup to series selection for a standard "large" test dataset, or to take no more than 2 seconds to go from series selection to display of the first image.
  
''<If the user might have an appreciation for the problems addressed in the profile, you can mention them here, but keep it short.  E.g. Mapping HL7 Order fields to DICOM Modality Worklist attributes can be inconsistent in the marketplace, so Scheduled Workflow provides vendors with more detailed instructions.>''
 
 
 
==Systems Affected==
 
==Systems Affected==
''<List (in user terms) the types of systems they might expect to have implemented actors from this profile, e.g. RIS, PACS, HIS, CAD Workstation, etc. and for each, how it would participate.>''
 
  
* ''PACS systems may store, manage, and/or display Evidence Documents.''
+
* Simple '''Display Workstations''' or Viewing Software might comply with BIR.
* ''Display systems may query, retrieve and display Evidence Documents.''
+
* '''CD Burning''' Workstations might offer to put a BIR compliant viewer on the media.
* ''Reporting workstations may retrieve, process and include details from Evidence Documents in reports
+
* PACS systems might provide a '''web portal''' that complies with BIR.
 +
* '''EHR''' systems might provide image viewing that complies with BIR.
 +
 
  
 
'''Actors & Transactions:'''
 
'''Actors & Transactions:'''
Line 57: Line 80:
 
==Specification==
 
==Specification==
  
'''Profile Status:''' [[Comments| Final Text]]   
+
'''Profile Status:''' [[Comments| Trial Implementation]]   
''<Replace "Final Text" with "Trial Implementation" or "Public Comment" as appropriate.>''
 
  
 
'''Documents:'''  
 
'''Documents:'''  
  
''<Provide direct links to the specific volumes or supplements, and list the volume sections relevant to this profile.  This is a simple inventory of official normative and informative text.  If you would like to provide a reading guide or walkthrough of what is in each of the different sections for implementers or users, do that in the Profile FAQ or the Profile Implementation Page linked below.  If the profile uses transactions from multiple Tech. Frameworks, repeat the structure below.>''
+
[http://www.ihe.net/Technical_Framework/upload/IHE_RAD_Suppl_BIR.pdf Basic Image Review Supplement]
 
 
[http://www.ihe.net/Technical_Framework/index.cfm#radiology IHE Radiology Technical Framework:]
 
:* [http://www.ihe.net/Technical_Framework/upload/ihe_tf_rev8.pdf Vol. 1] - Section 5 (SWF Profile)
 
:* [http://www.ihe.net/Technical_Framework/upload/ihe_tf_rev8-2.pdf Vol. 2] - Sections 4.8 to 4.10, 4.14 to 4.19, and 4.23
 
:* [http://www.ihe.net/Technical_Framework/upload/ihe_tf_rev8-3.pdf Vol. 3] - Appendix E
 
  
 
'''Underlying Standards:'''
 
'''Underlying Standards:'''
  
''<list all the standards on which the profile is based; if possible with links to sources>''
 
 
:* [http://dicom.nema.org DICOM]
 
:* [http://dicom.nema.org DICOM]
:* [http://www.hl7.org HL7]
 
:* ...
 
  
 
==See Also==
 
==See Also==
 
''<The following sections can be left out if there is nothing to point to.  This is just to show where such information can go.>''
 
 
  
 
'''Related Profiles'''
 
'''Related Profiles'''
 +
* [[Portable Data for Imaging]] [PDI] includes a Basic Image Review option for putting a BIR Compliant viewer on the media.
 +
* [[Cross-enterprise Document Sharing for Imaging]] [XDS-I] can be used to access images which could be displayed by a BIR viewer.
  
''<List profiles this one depends on, profiles that depend on this one, profiles that are synergistic with this one.  Start with the name of the other profile as a link and then explain the relationship.>''
 
  
* ''[[Reporting Workflow]] [RWF] may use Evidence Documents as inputs to the reporting process.''
+
Basic Image Review does not address domain specific needs such as those described in:
* ''[[Simple Image & Numeric Reports]] [SINR] may include data copied from Evidence Documents.''
+
* [[Nuclear Medicine Image]] specifies how Nuclear Medicine images and result screens are created, exchanged, used and displayed.
* ''[[Cross-enterprise Document Sharing for Imaging]] [XDS-I] can be used to share Evidence Documents between sites over a network.''
+
* [[Mammography Image]] specifies how Mammography images and evidence objects are created, exchanged, used and displayed.
* ''[[Portable Data for Imaging]] [PDI] can store Evidence Documents on media such as CDs.''
 
* ''[[Import Reconciliation Workflow]] [IRWF] can fix patient ids, etc. of Evidence Documents when importing.''
 
  
  
 
'''Consumer Information'''
 
'''Consumer Information'''
  
The [[{{{3|Profile FAQ Template}}}]] answers typical questions about what the Profile does.  ''<Replace the link with a link to the actual FAQ page for the Profile>''
+
The [[Basic Image Review FAQ]] answers typical questions about what the Profile does.   
  
The [[{{{3|Profile Purchasing Template}}}]] describes considerations when purchasing equipment to deploy this Profile.  ''<Replace the link with a link to the actual Purchasing page for the Profile>''
+
[[Basic Image Review Purchasing]] describes considerations when purchasing equipment to deploy this Profile.   
  
 
'''Implementer Information'''
 
'''Implementer Information'''
  
The [[{{{3|Basic Image Review Implementation}}}]] provides additional information about implementing this Profile in software.  ''<Replace the link with a link to the actual Implementation page for the Profile>''
+
The [[Basic Image Review Implementation]] provides additional information about implementing this Profile in software.   
  
 
'''Reference Articles'''
 
'''Reference Articles'''
  
''<List References (good and bad) (with link if possible) to Journal Articles that mention IHE's work (and hopefully include some analysis).  Go ahead, Google: IHE <Profile Name> abstract  or something.  You might be surprised. >''
 
  
  
Line 111: Line 120:
  
 
[[Category:Profiles]]
 
[[Category:Profiles]]
 
 
<noinclude>''<'''Delete this Category Templates line''' since your Profile page is no longer a template.>'' </noinclude>
 
[[Category:{{{4|Templates}}}]]
 

Revision as of 10:30, 23 July 2013

Basic Image Review (BIR) defines a set of baseline features and user interface relevant to the simple review of DICOM images.

Summary

Compliant software must provide a predictable user interface and functionality sufficient to review images for the purpose of clinical decision-making by ordering physicians:

  • display of grayscale and color images from any modality
  • visual navigation of the available series of images through the use of thumbnails
  • side-by-side comparison of at least two sets of images (with synchronized scroll, pan and zoom for cross-sectional modalities)
  • annotation of laterality, orientation, and spatial localization
  • annotation of demographics, management and basic technique information for safe identification and usage
  • simple measurements of linear distance and angle
  • cine capability for images that involve cardiac motion (e.g., cardiac US, XA, 500 CT or MR)

BIR is not intended to be a reading workstation.

Benefits

Reduce Errors or Delays in Patient Care

  • Reduce the need to work with unfamiliar interfaces
  • Reduce CD data that is effectively inaccessible by making sure viewers are runnable and usable
  • Reduce interpretation problems by ensuring critical basic functions are present

Improve Efficiency

  • Reduce time spent learning new viewer interfaces by requiring similar layout, icons and interactions
  • Reduce time wasted on viewers that do not run and images that do not load by requiring support for a basic reference platform and minimum installation footprint.

Reduce Cost and Dose

  • Avoid repeating scans by making the original results properly accessible

Details

To ensure adequate, predictable functionality, the profile defines a specific, simplified set of functions.

To minimize errors and learning time, the user interface elements are described to the level needed for consistency, while allowing the implementer some flexibility with the fine details.

The BIR user interface includes:

  • a strip or panel containing a thumbnail for each series
  • one or more viewports which open to display selected series following predictable default image guidance
  • a toolbar or panel containing a standard set of icons for the specified functions
    • patient selection
    • image, series, study navigation within a patient
    • viewport and tile layout control
    • viewport selection and synchronization control
    • windowing and greyscale inversion control and presets
    • scrolling, rotation, panning and zooming control
    • annotation, localizer and crosshair display control
    • distance and angle measurement
    • cine control
    • report display control
    • printing

The profile allows implementers to include an "Advanced" button which gives users the choice of moving past the initial familiar interface and gives vendors a place to demonstrate their creativity.

The profile does not require (or permit) advanced image processing like pseudo-coloring, multi-planar reconstruction, multi-modality fusion, surface rendering, volume rendering, etc. in the basic interface. Vendors may choose to provide such functionality in their Advanced interface.

An implementation may choose to support screen calibration such as that described in the Consistent Presentation of Images Profile however the profile does not mandate it since this often depends on hardware which may not be present on the system where the BIR software is being run. As a result, the software may inform the user that the display environment may or may not be adequate for diagnostic purposes. However to comply with the profile there shall be nothing about the software or the data which precludes diagnostic use so the software may NOT declare either to be unsuitable.

Finally, the profile does not address application or modality specific requirements, such as may be needed for primary interpretation of mammography by radiologists, or primary interpretation of nuclear medicine image, which are addressed by other profiles.

A Basic Review Option in Portable Data for Imaging specifies media creators be capable of including a BIR compliant viewer.

To address installation issues, BIR viewers on PDI media are required to be able to run on a basic reference platform and install with a minimum footprint.

To address performance issues, BIR viewers on PDI media are required to meet certain performance targets on the reference platform, e.g. take no more than 60 seconds to go from startup to series selection for a standard "large" test dataset, or to take no more than 2 seconds to go from series selection to display of the first image.

Systems Affected

  • Simple Display Workstations or Viewing Software might comply with BIR.
  • CD Burning Workstations might offer to put a BIR compliant viewer on the media.
  • PACS systems might provide a web portal that complies with BIR.
  • EHR systems might provide image viewing that complies with BIR.


Actors & Transactions:

<Insert an actor-transaction diagram, and or list of Content Definitions>

Specification

Profile Status: Trial Implementation

Documents:

Basic Image Review Supplement

Underlying Standards:

See Also

Related Profiles


Basic Image Review does not address domain specific needs such as those described in:

  • Nuclear Medicine Image specifies how Nuclear Medicine images and result screens are created, exchanged, used and displayed.
  • Mammography Image specifies how Mammography images and evidence objects are created, exchanged, used and displayed.


Consumer Information

The Basic Image Review FAQ answers typical questions about what the Profile does.

Basic Image Review Purchasing describes considerations when purchasing equipment to deploy this Profile.

Implementer Information

The Basic Image Review Implementation provides additional information about implementing this Profile in software.

Reference Articles



This page is based on the Profile Overview Template